Related Experiment Video
Updated: Jun 26, 2025

Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Insights into modifiers effects in differential mobility spectrometry: A data science approach for metabolomics and
Stepan Stepanovic1,2, Lysi Ekmekciu1, Bandar Alghanem1,3
1Life Sciences Mass Spectrometry, Department of Inorganic and Analytical Chemistry, University of Geneva, Geneva, Switzerland.
Abstract:
Utilizing a data-driven approach, this study investigates modifier effects on compensation voltage in differential mobility spectrometry-mass spectrometry (DMS-MS) for metabolites and peptides. Our analysis uncovers specific factors causing signal suppression in small molecules and pinpoints both signal suppression mechanisms and the analytes involved. In peptides, machine learning models discern a relationship between molecular weight, topological polar surface area, peptide charge, and proton transfer-induced signal suppression. The models exhibit robust performance, offering valuable insights for the application of DMS to metabolites and tryptic peptides analysis by DMS-MS.