Related Experiment Video
Updated: May 8, 2026

Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
Leaflet modification before transcatheter aortic valve implantation in patients at risk for coronary obstruction: the
Danny Dvir1, Didier Tchétché2, Martin B Leon3
1Department of Cardiology, Shaare Zedek Medical Center and Faculty of Medicine, Hebrew University of Jerusalem, P.O. Box 3235, Shmuel Bait 12 Street, Jerusalem 9103102, Israel.
Background And Aims:
This trial sought to assess the safety and efficacy of ShortCut, the first dedicated leaflet modification device, prior to transcatheter aortic valve implantation (TAVI) in patients at risk for coronary artery obstruction.
Methods:
This pivotal prospective study enrolled patients with failed bioprosthetic aortic valves scheduled to undergo TAVI and were at risk for coronary artery obstruction. The primary safety endpoint was procedure-related mortality or stroke at discharge or 7 days, and the primary efficacy endpoint was per-patient leaflet splitting success. Independent angiographic, echocardiographic, and computed tomography core laboratories assessed all images. Safety events were adjudicated by a clinical events committee and data safety monitoring board.
Results:
Sixty eligible patients were treated (77.0 ± 9.6 years, 70% female, 96.7% failed surgical bioprosthetic valves, 63.3% single splitting and 36.7% dual splitting) at 22 clinical sites. Successful leaflet splitting was achieved in all [100%; 95% confidence interval (CI) 94%-100.0%, P < .001] patients. Procedure time, including imaging confirmation of leaflet splitting, was 30.6 ± 17.9 min. Freedom from the primary safety endpoint was achieved in 59 [98.3%; 95% CI (91.1%-100%)] patients, with no mortality and one (1.7%) disabling stroke. At 30 days, freedom from coronary obstruction was 95% (95% CI 86.1%-99.0%). Within 90 days, freedom from mortality was 95% [95% CI (86.1%-99.0%)], without any cardiovascular deaths.
Conclusions:
Modification of failed bioprosthetic aortic valve leaflets using ShortCut was safe, achieved successful leaflet splitting in all patients, and was associated with favourable clinical outcomes in patients at risk for coronary obstruction undergoing TAVI.
Related Concept Videos
Aortic Regurgitation III: Medical Management
Peripheral Artery Disease III: Interprofessional Care
Peripheral Artery Disease V: Postoperative Nursing Management