Related Experiment Video
Updated: Jun 26, 2025

Procedure for Human Saphenous Veins Ex Vivo Perfusion and External Reinforcement
Published on: October 1, 2014
Hemodynamics of the VenusP Valve System™-an in vitro study
Huang Chen1,2, Milad Samaee1, Michael Tree3
1Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, United States.
Abstract:
This study aims to evaluate the fluid dynamic characteristics of the VenusP Valve System™ under varying cardiac outputs in vitro. A thorough hemodynamic study of the valve under physiological cardiac conditions was conducted and served as an independent assessment of the performance of the valve. Flow fields downstream of the valve near the pulmonary bifurcation were quantitatively studied by two-dimensional Particle Image Velocimetry (PIV). The obtained flow field was analyzed for potential regions of flow stasis and recirculation, and elevated shear stress and turbulence. High-speed en face imaging capturing the leaflet motion provided data for leaflet kinematic modeling. The experimental conditions for PIV studies were in accordance with ISO 5840-1:2021 standard, and two valves with different lengths and different orientations were studied. Results show good hemodynamics performance for the tested valves according to ISO 5840 standard without significant regions of flow stasis. Observed shear stress values are all well below established hemolysis limits.
More Related Videos
Related Concept Videos
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...
Veins as Blood Reservoirs