Commissioning and Validation of CT Number to SPR Calibration in Carbon Ion Therapy Facility
Yuya Miyasaka1, Takayuki Kanai2, Hikaru Souda1
1Department of Heavy Particle Medical Science, Yamagata University Graduate School of Medical Science, Yamagata, Japan.
Purpose:
We performed computed tomography (CT)-stopping power ratio (SPR) calibration in a carbon-ion therapy facility and evaluated SPR estimation accuracy.
Materials And Methods:
A polybinary tissue model method was used for the calibration of CT numbers and SPR. As a verification by dose calculation, we created a virtual phantom to which the CT-SPR calibration table was applied. Then, SPR was calculated from the change in the range of the treatment planning beam when changing to 19 different CT numbers, and the accuracy of the treatment planning system (TPS) calculation of SPR values from the CT-SPR calibration table was validated. As a verification by measurement, 5 materials (water, milk, olive oil, ethanol, 40% K2HPO4) were placed in a container, and the SPR was obtained by measurement from the change in the range of the beam that passed through the materials.
Results:
The results of the dose calculations of the TPS showed that the results agreed within 1% for the lower CT numbers up to 1000 HU, but there was a difference of 3.0% in the higher CT number volume. The difference between the SPR calculated by TPS and the SPR caused by the difference in the energy of the incident particles agreed within 0.51%. The accuracy of SPR estimation was measured, and the error was within 2% for all materials tested.
Conclusion:
These results indicate that the SPR estimation errors are within the range of errors that can be expected in particle therapy. From commissioning and verification results, the CT-SPR calibration table obtained during this commissioning process is clinically applicable.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT