Result of the left bundle branch area pacing data collection registry, an international multicenter study of
Pugazhendhi Vijayaraman1, David Foo2, Tien Siang Eric Lim3
1Geisinger Wyoming Valley Medical Center, Geisinger Heart Institute, Wilkes-Barre, Pennsylvania, USA.
Introduction:
Left bundle branch area (LBBA) pacing (LBBAP) has been proposed as an alternative therapy option in patients indicated for cardiac pacing to treat bradycardia or heart failure. The aim of the study was to evaluate the safety and effectiveness of LBBAP in patients implanted with a Tendril 2088 stylet-driven lead.
Methods:
The international retrospective data collection registry included 11 sites from 5 countries globally. Patients with attempted implants of the Tendril lead in the LBBA were followed for at least 6 months post the implant attempt. The primary safety and efficacy endpoints were freedom from LBBAP lead-related serious adverse events and the composite of LBBA capture threshold of ≤2.0 V and R-wave amplitudes ≥5 mV (or ≥value at implant), respectively.
Results:
Of 221 patients with attempted implants of the Tendril 2088 lead in the LBBA, 91.4% (202/221) had successful implants for LBBAP. Regardless of the LBBAP implant success, all patients were followed for at least 6 months (8.7 ± 7.3 months). Baseline characteristics: 44% female, 84% ≥65 years old, 34% coronary artery disease, and 86% of primary indications for pacemaker implant. Both primary safety and effectiveness endpoints were met (freedom from LBBAP lead-related serious adverse device effects of 99.5% and electrical performance composite success rate of 93%). The capture thresholds in LBBAP at implant and 6 months were 0.8 ± 0.3 V@0.4 ± 0.1 ms and 0.8 ± 0.3 V@0.4 ± 0.1 ms. The rate of patients with capture threshold rise ≥1 V was 1.5% through 6 months. The R-wave amplitudes in LBBAP at implant and 6 months were 9.3 ± 3.2 mV and 10.6 ± 3.0 mV.
Conclusions:
This large multicenter study demonstrates that the stylet-driven Tendril™ STS 2088 lead is safe and effective for LBBAP with high success and low complication rates.
More Related Videos
10:05Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pulse
The pulse serves as a clinical...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...