Performance of Subcutaneous Continuous Glucose Monitoring in Adult Critically Ill Patients Receiving Vasopressor
Ola Friman1,2, Navid Soltani1,2, Marcus Lind3,4,5
1Department of Physiology and Pharmacology, Section of Anaesthesia and Intensive Care, Karolinska Institutet, Stockholm, Sweden.
None:
Background: Subcutaneous continuous glucose monitoring (CGM) may facilitate glucose control in the ICU. We aimed to assess the accuracy of CGM (Dexcom G6) against arterial blood glucose (ABG) in adult critically ill patients receiving intravenous insulin infusion and vasopressor therapy. We also aimed to assess feasibility and tolerability of CGM in this setting. Methods: We included ICU patients receiving mechanical ventilation, insulin, and vasopressor therapy. Numerical accuracy was assessed by the mean absolute relative difference (MARD), overall, across arterial glucose strata, over different noradrenaline equivalent infusion rates, and over time since CGM start. MARD <14% was considered acceptable. Clinical accuracy was assessed using Clarke Error Grid (CEG) analysis. Feasibility outcome included number and duration of interrupted sensor readings due to signal loss. Tolerability outcome included skin reactions related to sensor insertion or sensor adhesives. Results: We obtained 2946 paired samples from 40 patients (18 with type 2 diabetes) receiving a median (IQR) maximum noradrenaline equivalent infusion rate of 0.18 (0.08-0.33) µg/kg/min during CGM. Overall, MARD was 12.7% (95% CI 10.7-15.3), and 99.8% of CGM readings were within CEG zones A and B. MARD values ≥14% were observed when ABG was outside target range (6-10 mmol/L [108-180 mg/dL]) and with noradrenaline equivalent infusion rates above 0.10 µg/kg/min. Accuracy improved with time after CGM start, reaching MARD values <14% after 36 h. We observed four episodes of interrupted sensor readings due to signal loss, ranging from 5 to 20 min. We observed no skin reaction related to sensor insertion or sensor adhesives. Conclusions: In our ICU cohort of patients receiving vasopressor infusion, subcutaneous CGM demonstrated acceptable overall numerical and clinical accuracy. However, suboptimal accuracy may occur outside glucose ranges of 6-10 mmol/L (108-180 mg/dL), during higher dose vasopressor infusion, and during the first 36 h after CGM start.
More Related Videos
07:35Author Spotlight: Investigating the Blood Glucose Homeostasis in Murine Brain Using a Cost-Effective Hyperglycemic And Hypoglycemic Clamp Technique
Published on: January 26, 2024
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Hypoglycemia and Glucagon
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...