Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Computed Tomography01:10

Computed Tomography

4.5K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
4.5K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A Case of Silicone Allergy Induced by a Subcutaneous Implantable Cardioverter Defibrillator.

Journal of arrhythmia·2026
Same author

Impact of nasogastric administration using the simple suspension method on serum tirabrutinib levels: A case report.

American journal of health-system pharmacy : AJHP : official journal of the American Society of Health-System Pharmacists·2026
Same author

Long RP tachycardia with fusion QRS during ventricular overdrive pacing: Diagnostic challenge of a bystander nodoventricular pathway.

HeartRhythm case reports·2026
Same author

A Novel Risk Score for Predicting the Difficulty in Leadless Pacemaker Implantation.

Journal of cardiovascular electrophysiology·2026
Same author

[A case of severe liver injury associated with erythropoietic protoporphyria].

Nihon Shokakibyo Gakkai zasshi = The Japanese journal of gastro-enterology·2025
Same author

Protocol for a prospective, multicenter, randomized, controlled trial comparing pulsed field ablation vs. cryoballoon ablation in patients with persistent atrial fibrillation (PEACE trial).

International journal of cardiology. Heart & vasculature·2025

Related Experiment Video

Updated: Jun 25, 2025

Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
08:19

Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences

Published on: May 17, 2018

9.8K

[Effects of Subject Position on Metal Artifact Reduction of a Reverse Shoulder Prosthesis Using Computed Tomography].

Tetsuya Ijichi1, Nariaki Tabata1, Yuna Kawahara1

  • 1Department of Radiology, Fukuoka University Chikushi Hospital.

Nihon Hoshasen Gijutsu Gakkai Zasshi
|May 22, 2024
PubMed
Summary

Single energy metal artifact reduction (SEMAR) effectively reduces artifacts from reverse shoulder prostheses. However, artifact reduction may decrease as the subject moves off the CT scanner's center.

Keywords:
computed tomography (CT)metal artifactreverse shoulder prosthesissingle energy metal artifact reduction (SEMAR)subject position

More Related Videos

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
07:45

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report

Published on: August 4, 2022

3.4K
Reverse Total Shoulder Arthroplasty
10:10

Reverse Total Shoulder Arthroplasty

Published on: July 5, 2011

43.2K

Related Experiment Videos

Last Updated: Jun 25, 2025

Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
08:19

Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences

Published on: May 17, 2018

9.8K
The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
07:45

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report

Published on: August 4, 2022

3.4K
Reverse Total Shoulder Arthroplasty
10:10

Reverse Total Shoulder Arthroplasty

Published on: July 5, 2011

43.2K

Area of Science:

  • Medical Imaging
  • Radiology
  • Biomedical Engineering

Context:

  • Computed tomography (CT) is crucial for imaging orthopedic implants.
  • Metal artifacts can obscure details of prostheses, impacting diagnosis.
  • Single energy metal artifact reduction (SEMAR) is a software-based technique to mitigate these artifacts.

Purpose:

  • To evaluate the impact of patient positioning on SEMAR effectiveness for reverse shoulder prostheses.
  • To quantify artifact reduction across different positions using CT.

Summary:

  • A water phantom with a reverse shoulder prosthesis was scanned at varying positions (0-150 mm off-center).
  • Images were acquired with and without SEMAR, and artifact levels were assessed physically (artifact index) and visually.
  • SEMAR significantly reduced metal artifacts, but artifact levels increased as the phantom moved further from the scanner's center.

Impact:

  • SEMAR is a valuable tool for reducing metal artifacts in reverse shoulder prosthesis imaging.
  • Optimal patient positioning (on-center) is crucial for maximizing SEMAR's benefits.
  • Findings can inform imaging protocols to improve diagnostic accuracy for patients with these implants.