Digital phantom versus patient-specific radiation dosimetry in adult routine thorax CT examinations
Antonios E Papadakis1, Vassiliki Giannakaki1, John Stratakis1
1University Hospital of Heraklion, Medical Physics Department, Stavrakia, Heraklion, Crete, Greece.
Purpose:
The aim of this study was to compare the organ doses assessed through a digital phantom-based and a patient specific-based dosimetric tool in adult routine thorax computed tomography (CT) examinations with reference to physical dose measurements performed in anthropomorphic phantoms.
Methods:
Two Monte Carlo based dose calculation tools were used to assess organ doses in routine adult thorax CT examinations. These were a digital phantom-based dosimetry tool (NCICT, National Cancer Institute, USA) and a patient-specific individualized dosimetry tool (ImpactMC, CT Imaging GmbH, Germany). Digital phantoms and patients were classified in four groups according to their water equivalent diameter (Dw). Normalized to volume computed tomography dose index (CTDIvol), organ dose was assessed for lungs, esophagus, heart, breast, active bone marrow, and skin. Organ doses were compared to measurements performed using thermoluminescent detectors (TLDs) in two physical anthropomorphic phantoms that simulate the average adult individual as a male (Alderson Research Labs, USA) and as a female (ATOM Phantoms, USA).
Results:
The average percent difference of NCICT to TLD and ImpactMC to TLD dose measurements across all organs in both sexes was 13% and 6%, respectively. The average ± 1 standard deviation in dose values across all organs with NCICT, ImpactMC, and TLDs was ± 0.06 (mGy/mGy), ± 0.19 (mGy/mGy), and ± 0.13 (mGy/mGy), respectively. Organ doses decreased with increasing Dw in both NCICT and ImpactMC.
Conclusion:
Organ doses estimated with ImpactMC were in closer agreement to TLDs compared to NCICT. This may be attributed to the inherent property of ImpactMC methodology to generate phantoms that resemble the realistic anatomy of the examined patient as opposed to NCICT methodology that incorporates an anatomical discrepancy between phantoms and patients.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...