Fat quantification in dual-layer detector spectral CT: How to handle iron overload, varying tube voltage and
Isabel Molwitz1, Graeme Michael Campbell2, Tobias Knopp3,4
1Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Objectives:
Opposed to other spectral CT techniques, fat quantification in dual-layer detector CT (dlCT) has only recently been developed. The impact of concomitant iron overload and dlCT-specific protocol settings such as the dose right index (DRI), a measure of image noise and tube current, on dlCT fat quantification was unclear. Further, spectral information became newly available <120 kV. Therefore, this study's objective was to evaluate the impact of iron, changing tube voltage, and DRI on dlCT fat quantification.
Material And Methods:
Phantoms with 0 and 8mg/cm3 iron; 0 and 5mg/cm3 iodine; 0, 10, 20, 35, 50, and 100% fat and liver equivalent, respectively, were scanned with a dlCT (CT7500, Philips, the Netherlands) at 100kV/20DRI, 120kV/20DRI, 140kV/20DRI, and at 120kV/16DRI, 120kV/24DRI. Material decomposition was done for fat, liver, and iodine (A1); for fat, liver, and iron (A2); and for fat, liver, and combined reference values of iodine and iron (A3). All scans were analyzed with reference values from 120kV/20DRI. For statistics, the intraclass correlation coefficient (ICC) and Bland-Altman analyses were used.
Results:
In phantoms with iron and iodine, results were best for A3 with a mean deviation to phantom fat of 1.3±2.6% (ICC 0.999 [95%-confidence interval 0.996-1]). The standard approach A1 yielded a deviation of -2.5±3.0% (0.998[0.994-0.999]), A2 of 6.1±4.8% (0.991[0.974-0.997]). With A3 and changing tube voltage, the maximal difference between quantified fat and the phantom ground truth occurred at 100kV with 4.6±2.1%. Differences between scans were largest between 100kV and 140kV (2.0%[-7.1-11.2]). The maximal difference of changing DRI occurred between 16 and 24 DRI with 0.4%[-2.2-3.0].
Conclusion:
For dlCT fat quantification in the presence of iron, material decomposition with combined reference values for iodine and iron delivers the most accurate results. Tube voltage-specific calibration of reference values is advisable while the impact of the DRI on dlCT fat quantification is neglectable.
More Related Videos
07:59Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
Published on: September 7, 2018
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...