Is Conduction System Pacing a Valuable Alternative to Biventricular Pacing for Cardiac Resynchronization Therapy?
Davide Castagno1,2, Francesco Zanon3, Gianni Pastore3
1Division of Cardiology, Department of Medical Sciences, University of Turin, 10126 Turin, Italy.
Abstract:
Cardiac resynchronization therapy (CRT) significantly improves clinical outcomes in patients with ventricular systolic dysfunction and dyssynchrony. Biventricular pacing (BVP) has a class IA recommendation for patients with symptomatic heart failure with reduced ejection fraction (HFrEF) and left bundle branch block (LBBB). However, approximately 30% of patients have a poor therapeutic response and do not achieve real clinical benefit. Pre-implant imaging, together with tailored programming and dedicated device algorithms, have been proposed as possible tools to improve success rate but have shown inconsistent results. Over the last few years, conduction system pacing (CSP) is becoming a real and attractive alternative to standard BVP as it can restore narrow QRS in patients with bundle branch block (BBB) by stimulating and recruiting the cardiac conduction system, thus ensuring true resynchronization. It includes His bundle pacing (HBP) and left bundle branch area pacing (LBBAP). Preliminary data coming from small single-center experiences are very promising and have laid the basis for currently ongoing randomized controlled trials comparing CSP with BVP. The purpose of this review is to delve into the emerging role of CSP as an alternative method of achieving CRT. After framing CSP in a historical perspective, the pathophysiological rationale and available clinical evidence will be examined, and crucial technical aspects will be discussed. Finally, evidence gaps and future perspectives on CSP as a technique of choice to deliver CRT will be summarized.
Related Concept Videos
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Electrophysiology of Normal Cardiac Rhythm
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Mechanism of Cardiac Arrhythmias
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and...