Relationship between oscillatory volume and oscillatory pressure in neonatal high-frequency oscillatory ventilators:
Emanuela Zannin1, Camilla Rigotti2, Raffaele L Dellacà3
1Neonatal Intensive Care Unit, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy.
Objective:
We analysed the relationship between oscillatory volume (VOSC) and pressure amplitude (ΔP) in six neonatal high-frequency oscillatory (HFO) ventilators and related it to (1) the accuracy of VOSC and ΔP measurements and (2) the maximal delivered ΔP.
Design:
In vitro study.
Setting:
Neonatal intensive care unit.
Interventions:
Ventilators tested were VN800 (Dräger), Servo-n (Maquet Getinge), SensorMedics 3100A (Vyaire Medical), Fabian HFOi (Vyaire Medical), SLE6000 (SLE UK) and Humming Vue (Metran). We changed various settings and mechanical characteristics of the test lung to mimic preterm and term conditions.
Main Outcome Measures:
For each condition, we measured VOSC and ΔP. We assessed the accuracy of the VOSC and ΔP measurements versus a reference measurement system using linear regression and Bland-Altman analysis. We evaluated the maximum delivered ΔP at different oscillatory frequencies.
Results:
We observed large variability between machines in the ΔP displayed at any target VOSC. Most ventilators over-read ΔP with errors up to 30 cmH2O or 60%. The error in the measurement of VOSC was up to ±2 mL or ±30%. We observed high variability in the accuracy of ΔP and VOSC measurements; the SLE6000 committed the lowest errors in ΔP measurements and the Fabian HFOi in VOSC. The maximum delivered ΔP varied depending on the ventilator, being maximal for the Humming Vue, followed by the SLE6000 and SensorMedics 3100A.
Conclusions:
The variability in the relationship between VOSC and ΔP among HFO ventilators is largely explained by the variable accuracy in ΔP and VOSC measurement. Different ventilators also exhibit important differences in the maximal generated ΔP.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Respiratory Volumes and Capacities