Related Experiment Video
Updated: Jun 11, 2026

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Embolization of Patent Foramen Ovale Closure Device-Rare Complication and Unique Management Approach
Mila Kovacevic1,2, Marko Atanaskovic1, Katarina Obradovic1
1Faculty of Medicine, University of Novi Sad, 21000 Novi Sad, Serbia.
Abstract:
Percutaneous closure of the patent foramen ovale (PFO) is generally regarded as a safe and effective procedure, indicated in patients with a prior PFO-associated stroke. While it is highly safe, rarely, it could be accompanied by a migration of the device, mainly caused by the interplay of a specific PFO morphology and inappropriate device sizing. Herein, we outline a seldom-observed complication of an unintentional detachment of the PFO closure device during implantation, leading to its migration into the abdominal aorta, and a unique management approach. Due to the inability to recapture the occluder with a snare, which is considered to be a mainstay of endovascular retrieval methods, two coronary guidewires were maneuvered through the mesh of the occluder and then captured with a snare proximally to the occluder. This innovative dual-wire-snare system was carefully pulled to the common femoral artery, a position deemed suitable for surgical extraction via arteriotomy, which was achieved successfully.