Image quality of opportunistic breast examinations in photon-counting computed tomography: A phantom study
1Division of X-Ray Imaging and CT, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany; Medical Faculty, Heidelberg University, Im Neuenheimer Feld 672, 69120 Heidelberg, Germany.
Purpose:
To compare the breast imaging performance of a clinical whole-body photon-counting CT (PCCT) to that of a dedicated breast CT (BCT) to determine the image quality of opportunistic breast examinations in clinical PCCT.
Materials And Methods:
To quantify image quality for breast cancer applications, acquisitions of a breast phantom including representations of calcifications, fibers, and masses were performed using a clinical PCCT and a dedicated BCT. When imaging with the PCCT, the phantom was also combined with a thorax phantom to simulate realistic patient positioning, while only the breast phantom was imaged in the BCT. Images in BCT were acquired at 7.0 mGy (CTDI16cm) and using 2.6 mGy-25.0 mGy in the PCCT. Spatial resolution between the BCT and PCCT images was matched and data were reconstructed using the default methods of each system. The dose-normalized contrast-to-noise ratio (CNRD) of masses and the structural visibility of fibers and calcifications were evaluated as figures of merit for all reconstructions.
Results:
CNRD between masses and background was 0.56 mGy-½, on average with BCT and varied between 0.39 mGy-½ to 1.46 mGy-½ with PCCT over all dose levels, phantom configurations, and reconstruction algorithms. Calcifications down to a size of 0.29 mm and fibers down to a size of 0.23 mm could be reliably identified in the images of both systems.
Conclusions:
Clinical PCCT provides an image quality superior to that obtained with BCT in terms of CNRD and allows for the identification of calcifications and fibers at comparable dose levels.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Ultrasonography
During an ultrasonography procedure, a handheld device called...