IVsight as an Infusion Monitor for Patients Receiving Intravenous Therapy: An Exploratory, Unblinded, Single-Center
Carlos Mejia-Chew1, Brett Heuring2, Jeffrey Salmons2
1Division of Infectious Diseases, Department of Medicine, Washington University School of Medicine in St. Louis, St. Louis, Missouri.
Background:
Intravenous (IV) infusion therapy is commonly used in health care settings. However, IV therapy at home can be challenging because it relies on the patient's ability to manage multiple medications correctly, which may lead to decreased treatment adherence.
Objective:
We aimed to assess the usability and accuracy of the IVsight monitor in capturing IV infusion data compared to manual recording.
Methods:
A prospective, single-center, usability study involving patients receiving IV infusion therapy at one of the BJC's Home Infusion suites was conducted to evaluate the accuracy, performance, and acceptability of the device IVsight as a monitor for IV infusions.
Results:
Of the 15 participants, the median (IQR) age was 46 years (36-55), 8 (53%) were female, and 13 (87%) were non-Hispanic white. Each participant received a median (IQR) of 4 (4-5) infusions during the study, and 68 infusions were observed overall. The intraclass correlation coefficient between the IVsight measurement and manual recording of infusion duration in seconds was excellent (ICC 0.97, 95% confidence interval 0.96-0.98). The Bland-Altman plot visually showed an acceptable limit of agreement for the 2 measurements, and the linear regression analysis revealed no significant proportional bias between the 2 methods for measuring the IV infusion time. None of the participants thought that IVsight was difficult to hold, use, clean, or store. Only one participant was concerned that the device could interfere with the IV infusion, and all participants felt comfortable with the device transmitting data to their health care providers.
Conclusions:
The IVsight infusion monitoring device showed near-perfect agreement on the recorded IV infusion duration compared with manual recording, and good acceptability among the study participants.
More Related Videos
08:44Continuous IV Infusion is the Choice Treatment Route for Arginine-vasopressin Receptor Blocker Conivaptan in Mice to Study Stroke-evoked Brain Edema
Published on: September 1, 2016
09:18Intraventricular Drug Delivery and Sampling for Pharmacokinetics and Pharmacodynamics Study
Published on: March 31, 2022
Related Concept Videos
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
Clinical Trials: Overview