Related Experiment Video
Updated: Aug 11, 2026

Rodent Working Heart Model for the Study of Myocardial Performance and Oxygen Consumption
Published on: August 16, 2016
What is the weakest point of a secured aortic cannula in central extracorporeal membrane oxygenation?
Kayla V Lucas1, Yuriy Stukov1, Matthew S Purlee1
1Congenital Heart Center, Division of Cardiovascular Surgery, Departments of Surgery and Pediatrics, University of Florida, Gainesville, Florida, USA.
Purpose:
The purpose of this study was to compare techniques for securing the aortic extracorporeal membrane oxygenation (ECMO) cannula, using in vitro models.
Methods:
Two models were studied: a tissue model using porcine aortas and a stand model replacing the aorta with a metal stand to study the system independent of the tissue. Interventions in each model were divided into three experimental groups: Group 1 (3-0 Prolene® + 20-French Medtronic Arterial Cannula EOPA™), Group 2 (4-0 Prolene® + 16-French Medtronic Arterial Cannula DLP Pediatric), and Group 3 (5-0 Prolene® + 8-French Medtronic Arterial Cannula DLP Pediatric). In separate experiments, both gradual and rapid forces were applied to the cannulas, starting with 9.8 Newtons and increasing exponentially if the cannula remained secured. Additionally, the method of securing the tourniquet and the number of ties securing the tourniquet to the cannula were evaluated.
Results:
In the tissue model, even with a minimum force of 9.8 Newtons, the suture pulled through the aortic tissue, leaving sutures and ties intact. In the stand model, two purse-string sutures secured by two ligaclips held the cannula reliably and withstood higher total force. Dislodgement was prevented at forces close to 60 Newtons with only two hemostatic clips included in cannulation.
Conclusions:
The weakest part of the aortic ECMO cannulation system using in vitro experiments was the tissue. Assuming that these experiments translate in vivo, it is therefore critical to prevent any pull on the cannulas by securing ECMO cannulas and ECMO tubing to both the patient and the patient's bed. Sutures with a larger diameter withstand more force. Two medium hemostatic clips can secure Prolene® sutures within snares as safely as a mosquito hemostat. Two polypropylene purse-string sutures secured by two hemostatic clips were most reliable at greater forces. The rationale for publishing our experiments in this manuscript is to (1) communicate our quantification of possible contributing factors to this rare and likely catastrophic complication of unintended decannulation, (2) increase awareness about this potential complication, and (3) increase vigilance to assure prevention of this dreaded complication.
More Related Videos
06:27Modifications of the Langendorff Method for Simultaneous Isolation of Atrial and Ventricular Myocytes from Adult Mice
Published on: May 13, 2021
09:54Lung Rapid Recovery Procurement Combined with Abdominal Normothermic Regional Perfusion in Controlled Donation after Circulatory Death
Published on: August 15, 2022
Related Concept Videos
Anatomy of the Circulatory System
Overview of the Vascular System
Structure of Blood Vessels
Arteries and Arterioles
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of blood...
Anatomy of Blood Vessels
Arteries
Arteries circulate oxygenated blood from the heart, except the pulmonary artery, which transports deoxygenated blood to the lungs. Large arteries, such as the aorta, have...