Individualized hemodialysis: Is similar hemodialysis adequacy possible using less water?
Ali Aktaş1, Hakan Özer2, İsmail Baloğlu3
1Department of Internal Medicine, Faculty of Meram Medicine, Necmettin Erbakan University, Konya, Turkiye.
Background And Aim:
There are over 60,000 hemodialysis (HD) patients in Türkiye, and the number of patients is increasing yearly. Dialysate flow rate (Qd) is a factor in HD adequacy. Approximately 150 L of water are consumed per session to prepare the dialysate. We aimed to investigate whether HD effectiveness can be achieved at a low Qd in different patient groups for the purpose of saving water.
Materials And Methods:
This prospective study included 81 HD patients from 2 centers. The patients underwent an aggregate total of 486 HD sessions, including 3 sessions at a Qd of 500 mL/min and 3 sessions at a Qd of 300 mL/min for each patient. We used online Kt/V readings recorded at the end of each dialysis session to compare the effectiveness of these 2 types of HD session performed at a different Qd.
Results:
The online Kt/V readings were similar between the standard (500) and low (300) Qd HD (1.51 ± 0.41 and 1.49 ± 0.44, respectively, p = 0.069). In the subgroup analyses, men had higher online Kt/V values at the standard Qd compared to the low Qd (1.35 ± 0.30 and 1.30 ± 0.32, respectively, p = 0.019), but the Kt/V values were not different for women. While the low Qd did not reduce online Kt/V in patients using small surface area dialysis membranes (1.75 ± 0.35 for 300 Qd and 1.75 ± 0.32 for 500 Qd, p = 0.931), it was associated with reduced online Kt/V in patients using large surface area dialysis membranes (1.12 ± 0.25 for 300 Qd and 1.17 ± 0.24 for 500 Qd, p = 0.006). The low Qd did not result in differences in online Kt/V among low-weight patients. However, online Kt/V values were better with the standard Qd in patients weighing 65 kg and above.
Conclusion:
In our study, dialysis adequacy at a reduced dialysate flow was not inferior for women, patients with low body weight, or patients using small surface area membranes. Individualized HD at a reduced Qd of 300 mL/min in eligible patients can save 48 L of water per HD session and an average of 7500 L of water per year.
Related Concept Videos
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Regulation of Water Output
Heart Failure Drugs: Diuretics
Formation of Dilute Urine
Filtrate Osmolarity in the PCT
Initially, as the filtrate passes through the proximal convoluted tubule (PCT), its...
Regulation of Water Intake
Renal Tubule and Collecting Duct
Proximal Convoluted Tubule (PCT):
The PCT is the initial segment of the renal tubule, extending from the Bowman's capsule that encloses the glomerulus. Its convoluted structure and microvilli-lined cells increase the surface area for reabsorption. The PCT reabsorbs glucose, amino acids, sodium, and water from the filtrate, ensuring essential...