Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Gas Chromatography: Overview of Detectors01:13

Gas Chromatography: Overview of Detectors

Detectors in gas chromatography (GC) help identify and quantify the components of a mixture by translating chemical properties into measurable signals, which are displayed on a chromatogram. Detectors can be categorized into two main types: destructive and non-destructive.
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II01:19

Gas Chromatography: Types of Detectors-II

In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
High-Performance Liquid Chromatography: Types of Detectors01:15

High-Performance Liquid Chromatography: Types of Detectors

The role of the detectors in High-Performance Liquid Chromatography (HPLC) is to analyze the solutes as they exit from the chromatographic column. The detector recognizes the solute's property and generates corresponding electrical signals, which are converted into a readable graph of the detector's response versus elution time called a chromatogram at the computer. There are several types of HPLC detectors, each with its own advantages and limitations, depending on the analyte properties and...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Direct measurement of three different deformations near the ground state in an atomic nucleus.

Communications physics·2025
Same author

Lung counting with a highly segmented HPGe detector.

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine·2024
Same author

Novel environmental monitoring detector for discriminating fallout and airborne radioactivity.

Scientific reports·2023
Same author

Technical note: Simulation of lung counting applications using Geant4.

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)·2023
Same author

An updated strategic research agenda for the integration of radioecology in the european radiation protection research.

Journal of environmental radioactivity·2021
Same author

Concentrations and number size distribution of fine and nanoparticles in a traditional Finnish bakery.

Journal of occupational and environmental hygiene·2017

Related Experiment Video

Updated: Jul 22, 2026

The Establishment of a Lung Colonization Assay for Circulating Tumor Cell Visualization in Lung Tissues
07:39

The Establishment of a Lung Colonization Assay for Circulating Tumor Cell Visualization in Lung Tissues

Published on: June 16, 2018

9.3K

Improving the detection limit in lung counting with a segmented HPGe detector.

Henri Jutila1, Paul Greenlees1, Tiina Torvela2

  • 1Accelerator Laboratory, Department of Physics, University of Jyväskylä, FI-40014 Jyväskylä, Finland; Helsinki Institute of Physics, University of Helsinki, P.O. Box 64, FI-00014 Helsinki, Finland.

Applied Radiation and Isotopes : Including Data, Instrumentation and Methods for Use in Agriculture, Industry and Medicine
|May 30, 2024
PubMed
Summary

A new segmented High-Purity Germanium detector with active shielding significantly reduces background noise in lung counting. This improves the minimum detectable activity (MDA) for both internal and external radioactive sources, enhancing field measurements.

Keywords:
Geant4 simulationLung countingMDASegmented HPGe detectorVoxel phantom

More Related Videos

Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
06:53

Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm

Published on: July 23, 2020

5.6K
Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
08:05

Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia

Published on: December 19, 2020

14.1K

Related Experiment Videos

Last Updated: Jul 22, 2026

The Establishment of a Lung Colonization Assay for Circulating Tumor Cell Visualization in Lung Tissues
07:39

The Establishment of a Lung Colonization Assay for Circulating Tumor Cell Visualization in Lung Tissues

Published on: June 16, 2018

9.3K
Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
06:53

Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm

Published on: July 23, 2020

5.6K
Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
08:05

Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia

Published on: December 19, 2020

14.1K

Area of Science:

  • Nuclear instrumentation and radiation detection.
  • Medical physics and computational modeling.
  • Environmental and occupational health physics.

Background:

  • Lung counting applications require sensitive detection of internal radioactive contamination.
  • Compton scattering from internal and external sources creates background noise in High-Purity Germanium (HPGe) detector spectra.
  • Minimizing background is crucial for accurate assessment of inhaled radionuclides like Americium-241 (Am-241).

Purpose of the Study:

  • To investigate the effectiveness of a segmented HPGe detector with active and passive shielding in reducing background events for lung counting.
  • To evaluate the impact of this configuration on the Minimum Detectable Activity (MDA) compared to traditional single-crystal HPGe detectors.
  • To assess the suitability of the proposed detector system for field measurements.

Main Methods:

  • Simulations were performed using the Geant4 toolkit.
  • Eight different segmented HPGe detector models with various shielding configurations were tested.
  • Inhaled Am-241 activity was simulated in the lungs of an ICRP adult reference computational phantom.
  • The reduction in MDA for internal and external sources was calculated and compared.

Main Results:

  • The segmented HPGe detector combined with active shielding demonstrated a significant reduction in background events.
  • A reduction in MDA of up to 30% for internal sources and 66% for external sources was achieved.
  • The proposed detector configuration effectively lowers the detection limit and/or measurement time.

Conclusions:

  • The segmented HPGe detector with active shielding is a promising advancement for lung counting applications.
  • This technology can improve the accuracy and efficiency of monitoring internal radioactive contamination.
  • The system is particularly advantageous for mobile or field-based radiation monitoring scenarios.