Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating
Published on: October 11, 2016
Optimization of contrast and dose in x-ray phase-contrast tomography with a Talbot-Lau interferometer
Henrik Mäkinen1, Heikki Suhonen1, Teemu Siiskonen1,2
1Department of Physics, PO Box 64, FI-00014 University of Helsinki, Finland.
Abstract:
X-ray phase-contrast imaging has become a valuable tool for biomedical research due to its improved contrast abilities over regular attenuation-based imaging. The recently emerged Talbot-Lau interferometer can provide quantitative attenuation, phase-contrast and dark-field image data, even with low-brilliance x-ray tube sources. Thus, it has become a valid option for clinical environments. In this study, we analyze the effects of x-ray tube voltage and total number of images on the contrast-to-noise ratio (CNR) and dose-weighted CNR (CNRD) calculated from tomographic transmission and phase-contrast data of a phantom sample. Constant counting statistics regardless of the voltage was ensured by adjusting the image exposure time for each voltage setting. The results indicate that the x-ray tube voltage has a clear effect on both image contrast and noise. This effect is amplified in the case of phase-contrast images, which is explained by the polychromatic x-ray spectrum and the dependence of interferometer visibility on the spectrum. CNRD is additionally affected by the total imaging time. While submerging the sample into a water container effectively reduces image artefacts and improves the CNR, the additional attenuation of the water must be compensated with a longer exposure time. This reduces dose efficiency. Both the CNR and CNRD are higher in the phase-contrast images compared to transmission images. For transmission images, and phase-contrast images without the water container, CNRD can be increased by using higher tube voltages (in combination with a lower exposure time). For phase-contrast images with the water container, CNRD is increased with lower tube voltages. In general, the CNRD does not strongly depend on the number of tomographic angles or phase steps used.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...