Related Experiment Video
Updated: Aug 2, 2026

Thermal Ablation for the Treatment of Abdominal Tumors
Published on: March 7, 2011
Assessment of stone ablation rate using the Moses technology modes with different energy and pulse settings: An
BegonaBallesta Martinez1, Panteleimon Ntasiotis2, Paraskevi Katsakiori2
1Department of Urology, University of Patras, Patras, Spain.
Objectives:
To compare lithotripsy ablation rate with the Moses modes versus conventional pulse modes when using the Holmium:Yttrium-Aluminum-Garnet (Ho:YAG) laser.
Methods:
The Lumenis® Pulse P120H Holmium Laser System and a 365 μm Moses D/F/L fiber were used to assess stone ablation rate in conventional Short and Long Pulse as well as Moses Contact and Distance at 10 W (0.5Jx20Hz and 2Jx5Hz) and 60 W (1Jx60Hz and 2Jx30Hz). Hard and soft phantom stones were formed, and all tests were conducted in a custom experimental configuration installed in a saline-filled bath. The laser was delivered up to 3 kJ of total energy. The fragmentation pattern was assessed via photographs in each cohort.
Results:
The time to reach the target energy was 5 min and 50 s in all 10 W and 60 W trials, respectively. In both stone types, ablation was more effective when high-power, high-energy and Moses Distance was utilized. In soft stones, the lowest ablation rate was detected in the Long Pulse modality in all power, energy and frequency settings. Overall, when dusting settings (high-frequency, low-energy) were used, a deeper single cavitation was observed rather than small cavitations.
Conclusions:
The most effective pulse modality as evaluated via stone ablation rate depends on the stone hardness as well as energy and frequency settings. In both hard and soft stones, ablation is more effective when 60 W (2Jx30Hz) power settings and Moses Distance are used. Tailored laser settings in terms of energy and frequency could be set for each case scenario.
More Related Videos
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
07:52Author Spotlight: Advancing Human Brain Modulation – Optimized Protocols for Transcranial Ultrasound Stimulation Experiments
Published on: June 28, 2024
Related Concept Videos
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Non-destructive Tests for Concrete Strength