Quantitative assessment of full-width at half-maximum and detector energy threshold in X-ray imaging systems
Abdollah Khorshidi1, Mansour Ashoor1
1Radiation Applications Research School, Nuclear Science and Technology Research Institute, Tehran, Iran.
Background:
The response function of imaging systems is regularly considered to improve the qualified maps in various fields. More the accuracy of this function, the higher the quality of the images.
Methods:
In this study, a distinct analytical relationship between full-width at half-maximum (FWHM) value and detector energy thresholds at distinct tube peak voltage of 100 kV has been addressed in X-ray imaging. The outcomes indicate that the behavior of the function is exponential. The relevant cut-off frequency and summation of point spread function S(PSF) were assessed at large and detailed energy ranges.
Results:
A compromise must be made between cut-off frequency and FWHM to determine the optimal model. By detailed energy range, the minimum and maximum of S(PSF) values were revealed at 20 keV and 48 keV, respectively, by 2979 and 3073. Although the maximum value of FWHM occurred at the energy of 48 keV by 224 mm, its minimum value was revealed at 62 keV by 217 mm. Generally, FWHM value converged to 220 mm and S(PSF) to 3026 with small fluctuations. Consequently, there is no need to increase the voltage of the X-ray tube after the energy threshold of 20 keV.
Conclusion:
The proposed FWHM function may be used in designing the setup of the imaging parameters in order to reduce the absorbed dose and obtain the final accurate maps using the related mathematical suggestions.
More Related Videos
06:51Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
Related Concept Videos
X-ray Imaging
Biological Effects of Radiation
Atomic Emission Spectroscopy: Overview