Related Experiment Video
Updated: Jul 25, 2026

09:17
High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation
Published on: July 29, 2011
14.8K
Aortic roadmapping during EVAR: a combined FEM-EM tracking feasibility study
Monica Emendi1, Geir A Tangen2, Pierluigi Di Giovanni3
1Department of Industrial Engineering, University of Tor Vergata, Rome, Italy. monica.emendi@alumni.uniroma2.eu.
Summary
This study introduces a novel approach using electromagnetic tracking and finite element modeling to improve intra-operative imaging during endovascular aneurysm repair (EVAR). This method enhances the accuracy of aortic roadmap predictions for safer navigation.
Area of Science:
- Biomedical Engineering
- Medical Imaging
- Computational Mechanics
Background:
- Intra-operative visualization in endovascular aneurysm repair (EVAR) currently relies on contrast-based imaging.
- Traditional image fusion techniques fail to account for dynamic vessel changes caused by guidewire insertion.
- Accurate real-time vessel mapping is crucial for safe and effective EVAR procedures.
Purpose of the Study:
- To develop and evaluate a novel image fusion approach combining electromagnetic (EM) tracking and finite element modeling (FEM).
- To improve intra-operative visualization by accounting for vessel deformations during EVAR.
- To enhance the accuracy of the predicted aortic roadmap for improved surgical navigation.
Main Methods:
- A patient-specific abdominal aorta model was created for segmentation and manufacturing.
- A database of numerical simulations with varying insertion angles was generated.
- An ad hoc tool with embedded EM sensors was designed for tracking during guidewire insertion.
- Cone beam computed tomography (CBCT) images were acquired to determine ground truth aortic deformations.
Main Results:
- The simulation that minimized the discrepancy between in silico predictions and in vitro sensor positions yielded the most accurate aortic displacement results.
- The integration of EM tracking data improved the predictive accuracy of the numerical simulations.
- Validation against CBCT-derived ground truth demonstrated the efficacy of the proposed method.
Conclusions:
- EM tracking technology can be utilized to minimize errors in predicted aortic roadmaps during EVAR.
- The combined EM tracking and FEM approach offers a promising solution for real-time, deformation-aware image fusion.
- This novel method has the potential to significantly enhance the safety and precision of EVAR procedures.