Design and validation of dual-point time-differentiated photoplethysmogram (2PPG) wearable for cuffless blood
Kei Fong Mark Wong1, Weiting Huang2, Dinaa Ying Hui Ee3
1School of Mechanical and Aerospace Engineering, Nanyang Technological University (NTU), Singapore.
Background & Objectives:
Measurement of blood pressure (BP) in ambulatory patients is crucial for at high-risk cardiovascular patients. A non-obtrusive, non-occluding device that continuously measures BP via photoplethysmography will enable long-term ambulatory assessment of BP. The aim of this study is to validate the metasense 2PPG cuffless wearable design for continuous BP estimation without ECG.
Methods:
A customized high-speed electronic optical sensor architecture with laterally spaced reflectance pulse oximetry was designed into a simple unobtrusive low-power wearable in the form of a watch. 78 volunteers with a mean age of 32.72 ± 7.4 years (21 to 64), 51% male, 49% female were recruited with ECG-2PPG signals acquired. The fiducial features of the 2PPG morphologies were then attributed to the estimator. A 9-1 K-fold cross-validation was applied in the ML.
Results:
The correlation for PTT-SBP was 0.971 and for PTT-DBP was 0.954. The mean absolute error was 3.167±1.636 mmHg for SBP and 6.4 ± 3.9 mm Hg for DBP. The ambulatory estimate for SBP and DBP for an individual over 3 days with 8-hour recordings was 0.70-0.81 for SBP and 0.42-0.51 for DBP with a ± 2.65 mmHg for SBP and ±2.02 mmHg for DBP. For SBP, 98% of metasense measurements were within 15 mm Hg and for DBP, 91% of metasense measurements were within 10 mmHg CONCLUSIONS: The metasense device provides continuous, non-invasive BP estimations that are comparable to ambulatory BP meters. The portability and unobtrusiveness of this device, as well as the ability to continuously measure BP could one day enable long-term ambulatory BP measurement for precision cardiovascular therapeutic regimens.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
06:16Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure