Improvement of Whole-body Bone Planar Images on a Bone-dedicated Single-photon Emission Computed Tomography Scanner
Zhexin Wang1,2, Hui Liu1,2, Li Cheng3
1Department of Engineering Physics, Tsinghua University, Beijing, China.
Purpose:
We have developed a bone-dedicated collimator with higher sensitivity but slightly degraded resolution on single-photon emission computed tomography (SPECT) for planar bone scintigraphy, compared with conventional low-energy high-resolution collimator. In this work, we investigated the feasibility of using the blind deconvolution algorithm to improve the resolution of planar images on bone scintigraphy.
Materials And Methods:
Monte Carlo simulation was performed with the NCAT phantom for modeling bone scintigraphy on the clinical dual-head SPECT scanner (Imagine NET 632, Beijing Novel Medical Equipment Ltd.) equipped with the bone-dedicated collimator. Maximum likelihood estimation method was used for the blind deconvolution algorithm. The initial estimation of point spread function (PSF) and iteration number for the method were determined by comparing the deblurred images obtained from different input parameters. We simulated different tumors in five different locations and with five different diameters to evaluate the robustness of the initial inputs. Furthermore, we performed chest phantom studies on the clinical SPECT scanner. The quantified increased contrast ratio (CR) between the tumor and the background was evaluated.
Results:
The 2 mm PSF kernel and 10 iterations provided a practical and robust deblurred image on our system. Those two inputs can generate robust deblurred images in terms of the tumor location and size with an average increased CR of 21.6%. The phantom studies also demonstrated the ability of blind deconvolution, using those two inputs, with increased CRs of 17%, 17%, 22%, 20%, and 13% for lesions with diameters of 1 cm, 2 cm, 3 cm, 4 cm, and 5 cm, respectively.
Conclusions:
It is feasible to use the blind deconvolution algorithm to deblur the planar images for SPECT bone scintigraphy. The appropriate values of the PSF kernel and the iteration number for the blind deconvolution can be determined using simulation studies.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:10A Sectioning, Coring, and Image Processing Guide for High-Throughput Cortical Bone Sample Procurement and Analysis for Synchrotron Micro-CT
Published on: June 12, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...