Magnum metal-on-metal uncemented total hip replacement: 8- to 18-year outcomes of 211 cases
D Gaillard-Campbell1, T Gross2
1Midlands Orthopaedics and Neurosurgery, PA, 1910 Blanding Street, Columbia, SC, 29201, USA. dani.gaillard@midorthoneuro.com.
Background:
Reports of adverse reactions to metal debris contributed in part to a decline in use of large-bearing metal-on-metal total hip devices. We hypothesize an optimal trunnion design may reduce risk of this failure mode in large-bearing total hip arthroplasty systems. The purpose of this study is to report mid- to long-term outcomes for a single-surgeon series of 211 total hip arthroplasties using the large-bearing Biomet Magnum metal-on-metal system.
Materials And Methods:
Between December 2004 and January 2016, the primary surgeon performed 211 uncemented Magnum total hip arthroplasties in 181 patients. The average length of follow-up was 10.1 ± 3.5 years (range 8-18 years).
Results:
Using failure of any component as the endpoint, the overall survivorship rate was 98.1% at 10 years and 97.4% at 18 years. These eight failures (3.8% of cohort) included one case of adverse wear-related failure (0.5%), two cases of acetabular ingrowth failure (0.9%), three cases of trunnion corrosion (1.4%), one failure of late infection (0.5%), and one inappropriate revision of components for trochanteric nonunion without instability (0.5%). Excluding failed cases, all components were radiographically stable with no radiolucencies. Except for the one wear failure, ion testing revealed that 97.2% of cases were within optimal whole blood metal ion levels with the remaining ion test results within acceptable levels.
Conclusions:
With the uncemented Magnum metal-on-metal total hip, we achieved 97.4% 18-year implant survivorship, exceeding the NICE criteria and registry benchmarks for implant survivorship. We observed a trunnion corrosion rate of 1.4% and no cases of instability. The single case of adverse wear-related failure was caused by acetabular component malposition.