Related Experiment Video
Updated: Jun 24, 2025

Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Evaluation of a Rebreathing System for Use with Portable Mechanical Ventilators
Introduction:
Maximizing the capabilities of available lowflow oxygen is key to providing adequate oxygen to prevent/treat hypoxemia and conserve oxygen. We designed a closed-circuit system that allows rebreathing of gases while scrubbing carbon dioxide (CO2) in conjunction with portable mechanical ventilators in a bench model.
Methods:
We evaluated the system using two portable mechanical ventilators currently deployed by the Department of Defense-Zoll 731 and AutoMedx SAVe II-over a range of ventilator settings and lung models, using 1 and 3L/min low-flow oxygen into a reservoir bag. We measured peak inspired oxygen concentration (FiO2), CO2-absorbent life, gas temperature and humidity, and the effect of airway suctioning and ventilator disconnection on FiO2 on ground and at altitude.
Results:
FiO2 was =0.9 across all ventilator settings and altitudes using both oxygen flows. CO2-absorbent life was >7 hours. Airway humidity range was 87%-97%. Mean airway temperature was 25.4°C (SD 0.5°C). Ten-second suctioning reduced FiO2 22%-48%. Thirtysecond ventilator disconnect reduced FiO2 29%-63% depending on oxygen flow used.
Conclusion:
Use of a rebreathing system with mechanical ventilation has the potential for oxygen conservation but requires diligent monitoring of inspired FiO2 and CO2 to avoid negative consequences.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...