An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
A Novel Bone-Screw-Fastener Demonstrates Greater Maximum Compression Force Before Failure Compared With a Traditional
Tyler Thorne1, Joseph Featherall, Dillon O'Neill
1Department of Orthopaedic Surgery, University of Utah, Salt Lake City, UT.
Objectives:
This study compared the maximal compression force before thread stripping of the novel bone-screw-fastener (BSF) with the traditional buttress screw (TBS) in synthetic osteoporotic and cadaveric bone models.
Methods:
The maximum compression force of the plate-bone interface before loss of screw purchase during screw tightening was measured between self-tapping 3.5-mm BSF and 3.5-mm TBS using calibrated load cells. Three synthetic biomechanical models were used: a synthetic osteoporotic diaphysis (model 1), a 3-layer biomechanical polyurethane foam with 50-10-50 pounds-per-cubic-foot layering (model 2), and a 3-layer polyurethane foam with 50-15-50 pounds-per-cubic-foot layering (model 3). For the cadaveric metaphyseal model, 3 sets of cadaveric tibial plafonds and 3 sets of cadaveric tibial plateaus were used. A plate with sensors between the bone and plate interface was used to measure compression force during screw tightening in the synthetic bone models, while an annular load cell that measured screw compression as it slid through a guide was used to measure compression in the cadaver models.
Results:
Across all synthetic osteoporotic bone models, the BSF demonstrated greater maximal compression force before stripping compared with the TBS [model 1, 155.51 N (SD = 7.77 N) versus 138.78 N (SD = 12.74 N), P = 0.036; model 2, 218.14 N (SD = 14.15 N) versus 110.23 N (SD = 8.00 N), P < 0.001; model 3, 382.72 N (SD = 20.15) versus 341.09 N (SD = 15.57 N), P = 0.003]. The BSF had greater maximal compression force for the overall cadaver trials, the tibial plafond trials, and the tibial plateau trials [overall, 111.27 N vs. 97.54 N (SD 32.32 N), P = 0.002; plafond, 149.6 N versus 132.92 N (SD 31.32 N), P = 0.006; plateau, 81.33 N versus 69.89 N (SD 33.38 N), P = 0.03].
Conclusions:
The novel bone-screw-fastener generated 11%-65% greater maximal compression force than the TBS in synthetic osteoporotic and cadaveric metaphyseal bone models. A greater compression force may increase construct stability, facilitate early weight-bearing, and reduce construct failure.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Wood Fasteners
Nails are the most common fasteners, consisting of sharp-pointed metal pins that are driven into wood using a hammer or a mechanical nail gun. They come in a variety of diameters, heads, and lengths. Nailing techniques include...
Normal Strain under Axial Loading
Relation Between Tensile Strength and Compressive Strength of Concrete
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen...