Cardiac Magnetic Resonance Imaging at 7 Tesla
Published on: January 6, 2019
Technical Note: Swing golden angle - A navigator-interleaved golden angle trajectory with eddy current suppression -
Zhongsen Li1, Aiqi Sun2, Chuyu Liu1
1Center for Biomedical Imaging Research, School of Biomedical Engineering, Tsinghua University, Beijing, China.
Background:
Golden angle (GA) radial trajectory is advantageous for dynamic magnetic resonance imaging (MRI). Recently, several advanced algorithms have been developed based on navigator-interleaved GA trajectory to realize free-running cardiac MRI. However, navigator-interleaved GA trajectory suffers from the eddy-current effect, which reduces the image quality.
Purpose:
This work aims to integrate the navigator-interleaved GA trajectory with clinical cardiac MRI acquisition, with the minimum eddy-current artifacts. The ultimate goal is to realize a high-quality free-running cardiac imaging technique.
Methods:
In this paper, we propose a new "swing golden angle" (swingGA) radial profile order. SwingGA samples the k-space by rotating back and forth at the generalized golden ratio interval, with smoothly interleaved navigator readouts. The sampling efficiency and angle increment distributions were investigated by numerical simulations. Static phantom imaging experiments were conducted to evaluate the eddy current effect, compared with cartesian, golden angle radial (GA), and tiny golden angle (tGA) trajectories. Furthermore, 12 heart-healthy subjects (aged 21-25 years) were recruited for free-running cardiac imaging with different sampling trajectories. Dynamic images were reconstructed by a low-rank subspace-constrained algorithm. The image quality was evaluated by signal-to-noise-ratio and spectrum analysis in the heart region, and compared with traditional clinical cardiac MRI images.
Results:
SwingGA pattern achieves the highest sampling efficiency (mSE > 0.925) and the minimum azimuthal angle increment (mAD < 1.05). SwingGA can effectively suppress eddy currents in static phantom images, with the lowest normalized root mean square error (nRMSE) values among radial trajectories. For the in-vivo cardiac images, swingGA enjoys the highest SNR both in the blood pool and myocardium, and contains the minimum level of high-frequency artifacts. The free-running cardiac images have good consistency with traditional clinical cardiac MRI, and the swingGA sampling pattern achieves the best image quality among all sampling patterns.
Conclusions:
The proposed swingGA sampling pattern can effectively improve the sampling efficiency and reduce the eddy currents for the navigator-interleaved GA sequence. SwingGA is a promising sampling pattern for free-running cardiac MRI.
More Related Videos
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI