Ambient light intensity affecting ultrasound operator detection of liver lesions in cine-clips
P Y C Lee1, A Mohamed Afif1, A Anthony1
1Department of Radiography, Allied Health Division, Singapore General Hospital, Singapore.
Introduction:
Ambient light (AL) is an important factor to improve ultrasound pathology detection. However, there are no established room AL levels recommended during an ultrasound examination. We aim to examine the diagnostic accuracy using different intensity of AL for the detection of liver lesions in anonymised pre-recorded cine-clips.
Methods:
Eight ultrasound operators with 5-14 years' professional experiences were prospectively recruited to evaluate 51 randomised cine-clips directly from one ultrasound machine. These 15-s clips of the right lobe of the liver in longitudinal and transverse planes were meant to simulate the ultrasound examination. Operators reviewed all cine-clips and responded to two questions per cine-clip regarding their detection performance under 3 AL settings; 3, 15 and 25 lux, at one lighting per visit. A repeat visit under each AL was performed to assess for intra-operator variability. Each operator completed six visits in total, with at least a 2-day washout period. The operators' performance was compared against imaging reference standards from contrast CT/MRI for cine-clips with lesion and serial US for those without.
Results:
AL with highest degree of diagnostic accuracy was found to be at 25 lux. Results from 8 operators revealed sensitivity ranged from 79% to 100%, specificity ranged from 94% to 100%. Positive and negative predictive values were up to 100% with AL at 25 lux. Both intra-and interrater reliability were excellent at 0.85-1.0 (0.79-0.98) and 0.98 (0.97, 0.99) respectively, with AL at 25 lux.
Conclusion:
This study proved that ambient light intensity affects the ultrasound operator detection of liver lesions on cine-clips.
Implications For Practice:
Identifying suitable AL levels will influence future ultrasound room construct.
More Related Videos
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
06:42Providing Visual Biofeedback Using Brightness Mode Ultrasound During a Golf Swing
Published on: August 25, 2022
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...