Validation of a Handheld 6-Lead Device for QT Interval Monitoring in Resource-Limited Settings
John Z Metcalfe1, Tamsin Economou2, Fahd Naufal1
1Division of Pulmonary and Critical Care Medicine, Zuckerberg San Francisco General Hospital and Trauma Center, University of California, San Francisco.
Importance:
Rifampin-resistant tuberculosis treatment regimens require electrocardiographic (ECG) monitoring due to the use of multiple QTc-prolonging agents. Formal 12-lead ECG devices represent a significant burden in resource-constrained clinics worldwide and a potential barrier to treatment scale-up in some settings.
Objective:
To evaluate the diagnostic accuracy of a handheld 6-lead ECG device within resource-constrained clinics.
Design, Setting, And Participants:
This diagnostic study was performed within a multicenter, pragmatic (broad eligibility criteria with no exclusions for randomized participants), phase 3 rifampin-resistant tuberculosis treatment trial (BEAT Tuberculosis [Building Evidence for Advancing New Treatment for Tuberculosis]) in South Africa. A total of 192 consecutive trial participants were assessed, and 191 were recruited for this substudy between January 21, 2021, and March 27, 2023. A low proportion (3 of 432 [0.7%]) of all screened trial participants were excluded due to a QTc interval greater than 450 milliseconds. Triplicate reference standard 12-lead ECG results were human calibrated with readers blinded to 6-lead ECG results.
Main Outcomes And Measures:
Diagnostic accuracy, repeatability, and feasibility of a 6-lead ECG device.
Results:
A total of 191 participants (median age, 36 years [IQR, 28-45 years]; 81 female participants [42.4%]; 91 participants [47.6%] living with HIV) with a median of 4 clinic visits (IQR, 3-4 visits) contributed 2070 and 2015 12-lead and 6-lead ECG assessments, respectively. Across 170 participants attending 489 total clinic visits where valid triplicate QTc measurements were available for both devices, the mean 12-lead QTc measurement was 418 milliseconds (range, 321-519 milliseconds), and the mean 6-lead QTc measurement was 422 milliseconds (range, 288-574 milliseconds; proportion of variation explained, R2 = 0.4; P < .001). At a QTc interval threshold of 500 milliseconds, the 6-lead ECG device had a negative predictive value of 99.8% (95% CI, 98.8%-99.9%) and a positive predictive value of 16.7% (95% CI, 0.4%-64.1%). The normal expected range of within-individual variability of the 6-lead ECG device was high (±50.2 milliseconds [coefficient of variation, 6.0%]) relative to the 12-lead ECG device (±22.0 milliseconds [coefficient of variation, 2.7%]). The mean (SD) increase in the 12-lead QTc measurement during treatment was 10.1 (25.8) milliseconds, with 0.8% of clinic visits (4 of 489) having a QTc interval of 500 milliseconds or more.
Conclusions And Relevance:
This study suggests that simplified, handheld 6-lead ECG devices are effective triage tests that could reduce the need to perform 12-lead ECG monitoring in resource-constrained settings.
More Related Videos
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...