Published on: September 7, 2018
Free-Breathing High-Resolution, Swap-Free, and Motion-Corrected Water/Fat Separation in Pediatric Abdominal MRI
Reyhaneh Nosrati1, Fatih Calakli, Onur Afacan
1From the Department of Radiology, Boston Children's Hospital, Boston, MA (R. Nosrati, F.C., O.A., K.P., R. Nichols, P.C., M.A.B., A.T., S.B., S.K.W.); and Harvard Medical School, Boston, MA (R. Nosrati, F.C., O.A., K.P., R. Nichols, P.C., M.A.B., A.T., S.B., S.K.W.).
Objectives:
The T1-weighted GRE (gradient recalled echo) sequence with the Dixon technique for water/fat separation is an essential component of abdominal MRI (magnetic resonance imaging), useful in detecting tumors and characterizing hemorrhage/fat content. Unfortunately, the current implementation of this sequence suffers from several problems: (1) low resolution to maintain high pixel bandwidth and minimize chemical shift; (2) image blurring due to respiratory motion; (3) water/fat swapping due to the natural ambiguity between fat and water peaks; and (4) off-resonance fat blurring due to the multipeak nature of the fat spectrum. The goal of this study was to evaluate the image quality of water/fat separation using a high-resolution 3-point Dixon golden angle radial acquisition with retrospective motion compensation and multipeak fat modeling in children undergoing abdominal MRI.
Materials And Methods:
Twenty-two pediatric patients (4.2 ± 2.3 years) underwent abdominal MRI on a 3 T scanner with routine abdominal protocol and with a 3-point Dixon radial-VIBE (volumetric interpolated breath-hold examination) sequence. Field maps were calculated using 3D graph-cut optimization followed by fat and water calculation from k-space data by iteratively solving an optimization problem. A 6-peak fat model was used to model chemical shifts in k-space. Residual respiratory motion was corrected through soft-gating by weighting each projection based on the estimated respiratory motion from the center of the k-space. Reconstructed images were reviewed by 3 pediatric radiologists on a PACS (picture archiving and communication systems) workstation. Subjective image quality and water/fat swapping artifact were scored by each pediatric radiologist using a 5-point Likert scale. The VoL (variance of Laplacian) of the reconstructed images was used to objectively quantify image sharpness.
Results:
Based on the overall Likert scores, the images generated using the described method were significantly superior to those reconstructed by the conventional 2-point Dixon technique ( P < 0.05). Water/fat swapping artifact was observed in 14 of 22 patients using 2-point Dixon, and this artifact was not present when using the proposed method. Image sharpness was significantly improved using the proposed framework.
Conclusions:
In smaller patients, a high-quality water/fat separation with sharp visualization of fine details is critical for diagnostic accuracy. High-resolution golden angle radial-VIBE 3-point Dixon acquisition with 6-peak fat model and soft-gated motion correction offers improved image quality at the expense of an additional ~1-minute acquisition time. Thus, this technique offers the potential to replace the conventional 2-point Dixon technique.
Insights
This study introduces an improved 3-point Dixon MRI technique for pediatric abdominal imaging, significantly enhancing water/fat separation and image sharpness. The new method overcomes limitations of conventional 2-point Dixon, offering superior diagnostic quality in children.
Area of Science:
- Medical Imaging
- Radiology
- Biophysics
Background:
- T1-weighted gradient recalled echo (GRE) sequences with Dixon technique are crucial for abdominal MRI, aiding tumor detection and tissue characterization.
- Current Dixon techniques face challenges including low resolution, motion blurring, water/fat swapping, and off-resonance artifacts.
Purpose of the Study:
- To evaluate the image quality of water/fat separation using a novel high-resolution 3-point Dixon golden angle radial acquisition.
- To assess the efficacy of retrospective motion compensation and multipeak fat modeling in pediatric abdominal MRI.
Main Methods:
- A 3-point Dixon radial-VIBE sequence was used in 22 pediatric patients on a 3 T scanner.
- Image reconstruction involved 3D graph-cut optimization for field maps and iterative k-space optimization with a 6-peak fat model.
- Respiratory motion was corrected using soft-gating, and image sharpness was objectively quantified using variance of Laplacian (VoL).
Main Results:
- The novel 3-point Dixon technique yielded significantly superior image quality compared to conventional 2-point Dixon (P < 0.05).
- Water/fat swapping artifacts, present in 14/22 patients with 2-point Dixon, were eliminated with the proposed method.
- Significant improvements in image sharpness were observed with the new framework.
Conclusions:
- High-quality water/fat separation and sharp visualization are critical for diagnostic accuracy in pediatric abdominal MRI.
- The high-resolution golden angle radial-VIBE 3-point Dixon technique with advanced modeling and motion correction improves image quality.
- This advanced technique shows potential to replace conventional 2-point Dixon MRI, despite a slight increase in acquisition time.