Diamedica Draw-over Vaporiser: bench testing the UK Defence Anaesthesia System in the deployed environment.
Sebastian Bourn1,2, O Rylah2,3, T Fishenden2
1Anaesthetics, Royal Infirmary of Edinburgh, Edinburgh, UK s.bourn@nhs.net.
BMJ Military Health
|June 11, 2024
Summary
The Diamedica Draw-over Vaporiser 2 (DDV2) reliably delivers sevoflurane in a modified Defence Anaesthesia System, with predictable output across various settings, minute volumes, and temperatures, supporting its use in deployed anesthesia.
Area of Science:
- Anesthesiology
- Medical Devices
- Engineering
Background:
- The Diamedica Draw-over Vaporiser 2 (DDV2) is crucial for deployed volatile general anesthesia in the UK Defence Medical Services.
- The Defence Anaesthesia System utilizes a modified 'push-over' configuration of the DDV2 with a turbine-driven ventilator.
Purpose of the Study:
- To investigate sevoflurane delivery from the DDV2 in a modified Defence Anaesthesia System configuration.
- To assess the impact of varying minute volumes (MVs), vaporiser settings, and ambient temperatures on sevoflurane output.
Main Methods:
- Tested DDV2 settings (1-4%, induction) and MVs (2-8 L/min) at 20°C and 30°C for 30 minutes.
- Conducted a supplemental 90-minute experiment simulating damage control surgery (2% setting, 6 L/min MV).
- Monitored sevoflurane delivery phases: 'wash-in' and 'maintenance'.
Main Results:
- Sevoflurane delivery exhibited a 'wash-in' phase (<5 min) followed by a 'maintenance period'.
- Output was constant at low MVs/settings; it decreased predictably at higher MVs/settings.
- At 20°C, delivery was within 20% of the set value; higher settings, MVs, and temperatures increased variation due to incomplete temperature compensation.
Conclusions:
- The DDV2 demonstrates predictable function across tested settings, MVs, and temperatures in the Defence Anaesthesia System.
- Sevoflurane delivery characteristics are comparable to standard draw-over configurations.
- The equipment proved reliable and robust, supporting its continued use for deployed anesthesia and training.