Concordance between platelet counts by impedance and optical technique during microcytic anemia: towards a threshold

Mayssa Ben Ayed1,2,3, Bochra Gharbi4,2, Eya Jaouachi4,2

  • 1La Rabta University Hospital Center, Hematology department, Tunis, Tunisia.

La Tunisie Medicale
|June 12, 2024
PubMed
Abstract

Insights

In microcytosis cases, impedance platelet counts may be inaccurate. Optical platelet counts are recommended for MCV < 72.9 fL or when schizocytes are present for reliable results.

Area of Science:

  • Hematology
  • Clinical Pathology
  • Laboratory Medicine

Background:

  • Platelet count is critical for clinical decisions.
  • Impedance-based platelet counts (PLT-I) can overestimate platelet levels in microcytic conditions.

Purpose of the Study:

  • Compare impedance (PLT-I) and optical (PLT-O) platelet counting techniques.
  • Determine a Mean Corpuscular Volume (MCV) threshold for utilizing PLT-O.

Main Methods:

  • Prospective analytical study over two months.
  • Collected blood samples in K2 EDTA tubes for complete blood count analysis.
  • Utilized Sysmex-XN1000 for PLT-O analysis in channel-Ret mode.
  • Recorded fragmented red cells (FRC%) and microcytic red cells (Micro-R%).
  • Examined blood smears for interfering particles.

Main Results:

  • Strong correlation between PLT-I and PLT-O (r=0.971 for all, r=0.90 for PLT <150 x 10^9/L).
  • Bland-Altman analysis showed a bias of 16.53 x 10^9/L.
  • An MCV cut-off of 72.9 fL demonstrated promising performance for recommending PLT-O.
  • Less favorable results were observed with Micro-R% analysis.

Conclusions:

  • Impedance platelet counting is suitable for routine use.
  • Employ PLT-O for validation when MCV < 72.9 fL or schizocytes are detected.
  • Optical method enhances accuracy in specific microcytic scenarios.