A leadless ventricular pacemaker providing atrioventricular synchronous pacing in the real-world setting: 12-Month
Christophe Garweg1, Jason S Chinitz2, Eloi Marijon3
1University Hospitals Leuven, Leuven, Belgium.
Background:
Advances in leadless pacemaker technology have enabled accelerometer-based atrioventricular (AV) synchronous pacing by sensing atrial mechanical contraction.
Objective:
The purpose of this study was to report the performance of the Micra AV leadless pacemaker from the worldwide Micra AV post-approval registry (PAR) through 12 months.
Methods:
The Micra AV PAR is a prospective, single-arm, observational registry designed to assess the safety and effectiveness of Micra AV in a real-world setting. For the present interim analysis, major complications and system revisions through 12 months were summarized and compared to a historical cohort of 2667 patients implanted with a transvenous dual-chamber pacing system.
Results:
The device was successfully implanted in 796 of 801 patients (99.4%) at 97 centers in 19 countries. Micra AV patients were older (74.1 years vs 71.1 years; P < .0001) with a higher incidence of renal disease (22.3% vs 9.8%; P < .0001) compared with transvenous dual-chamber patients. Through 12 months, the major complication rate was 3.7% in Micra AV patients compared with 8.8% in transvenous dual-chamber patients (hazard ratio 0.42; 95% confidence interval 0.28-0.61; P < .001). The system revision rate was 1.5% in Micra AV patients compared with 5.5% for transvenous dual-chamber patients (hazard ratio 0.25; 95% confidence interval 0.13-0.47; P < .001); this reduction was largely driven by the absence of lead dislodgments requiring revision. The median AV synchrony index was 79.4% (interquartile range 65.2%-86.4%) in patients paced >90%.
Conclusion:
The Micra AV leadless pacemaker was implanted with a high rate of success in patients with multiple comorbidities, with a significantly lower rate of complications and system revisions through 12 months compared to a historical cohort of patients with transvenous dual-chamber pacemakers.
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Mechanism of Cardiac Arrhythmias
Heart Failure Drugs: Inotropic Agents
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials