MISSTEC-S: A fast 1H pulse calibration from spectra simultaneously produced by a spin echo and a stimulated echo
Margot Sanchez1, Julien Pontabry2, Gaëtan Assemat2
1CEISAM, Interdisciplinary Chemistry: Synthesis, Analysis, Modeling, Nantes University-CNRS UMR 6230, 2 rue de la Houssinière, BP 92208, F-44322 Nantes Cedex 3, France; RS(2)D, 13 rue Vauban, F-67450 Mundolsheim, France.
Abstract:
Radio-Frequency (RF) pulse calibration is an essential step in guaranteeing both optimum acquisition quality in multi-pulse NMR and accurate results in quantitative experiments. Most existing methods are based on a series of spectra for which the flip angle of one or more pulses is progressively incremented, implying a significant experiment time. In order to circumvent this drawback, we have previously proposed an approach based on the acquisition of a spin echo and a stimulated echo - the MISSTEC sequence - which requires only 8 s to determine the PW90-1H, while it is several minutes in the case of the use of a nutation curve. In this work, a new sequence for RF calibration is presented: MISSTEC-S. It is derived from the previously proposed MISSTEC sequence, but the observation of echoes in presence of magnetic field gradient is replaced by the observation of FIDs. This modification allows both spectra to be phased, while imposing a strong constraint on the Mixing Time (TM). However, the relationship used to calculate the flip angle is only correct when TM is small enough to neglect longitudinal relaxation during this delay. In order to reduce TM, the first FID is truncated during acquisition and subsequently lengthened using points from the second FID. Results obtained with MISSTEC-S were compared to those obtained from a complete nutation curve and an excellent correlation was observed, although the experimental time to obtain the PW90 is dramatically reduced.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
NMR Spectrometers: Resolution and Error Correction
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
NMR Spectroscopy: Spin–Spin Coupling