Evaluation of Noise Exposure Levels in Pediatric ENT Operating Rooms
Leyn Shakhtour1, Sophia Song1, Nicklas C Orobello2
1Division of Otolaryngology-Head and Neck Surgery, The George Washington University School of Medicine and Health Sciences, Washington, District of Columbia, USA.
Objective:
Operating room (OR) sounds may surpass noise exposure thresholds and induce hearing loss. Noise intensity emitted by various surgical instruments during common pediatric otolaryngologic procedures were compared at the ear-level of the surgeon and patient to evaluate the need for quality improvement measures.
Study Design:
Cross-sectional study.
Setting:
Single tertiary care center.
Methods:
Noise levels were measured using the RISEPRO Sound Level Meter and SoundMeter X 10.0.4 at the ear level of surgeon and patient every 5 minutes. Operative procedure and instrument type were recorded. Measured noise levels were compared against ambient noise levels and the Apple Watch Noise application.
Results:
Two hundred forty-two total occasions of noise were recorded across 62 surgical cases. Cochlear implantation surgery produces the loudest case at the ear-level of the patient (91.8 Lq Peak dB; P < .001). The otologic drill was the loudest instrument for the patient (92.1 Lq Peak dB; P < .001), while the powered microdebrider was the loudest instrument for the surgeon (90.7 Lq Peak dB; P = .036). Noise measurements between surgeon and patient were similar (P < .05). Overall agreement between the Noise application and Sound Level Meter was excellent (intraclass correlation coefficient of 0.8, with a 95% confidence interval ranging from 0.32 to 0.92).
Conclusion:
Otolaryngology OR noises can surpass normal safe thresholds. Failure to be aware of this may unwittingly expose providers to noise-related hearing loss. Mitigation strategies should be employed. Quality improvement measures, including attention to surgical instrument volume settings and periodic decibel measurements with sound applications, can promote long-term hearing conservation.
Discussion:
Otolaryngology OR noises can surpass normal safe thresholds. Failure to be aware of this may unwittingly expose providers to noise-related hearing loss. The duration, frequency of exposure, and volume levels of noise should be studied further.
Implications For Practice:
Mitigation strategies should be employed. Quality improvement measures, including attention to surgical instrument volume settings and periodic decibel measurements with sound applications, can promote long-term hearing conservation.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities:
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...