A Comparison of Two Transport Monitor Systems With Regard to Efficiency and Staff Satisfaction in the Perioperative
Kevin M Tourelle1, Jonas Fricke1, Manuel Feißt2
1Department of Anesthesiology, Medical Faculty Heidelberg, Heidelberg University, Heidelberg, DEU.
Background:
Medical research aims to improve patient safety and efficiency in the perioperative setting. One critical aspect of patient safety is the intrahospital transfer of patients. Also, reliable monitoring of vital signs is crucial to support the medical staff. This study was conducted to assess two monitoring systems in terms of the handover time and staff satisfaction.
Methods:
To assess several aspects, two monitoring systems were compared: an organizational unit-related monitoring system that needs to be changed and brought back to the initial organizational unit after the patient transfer and a patient-specific monitoring system that accompanies the patient during the whole perioperative process.
Results:
In total, 243 patients were included, and 375 transfers were examined to analyze economic factors, including differences in handover times and user-friendliness. To this end, 30 employees of the Heidelberg University Hospital were asked about their satisfaction with the two monitoring systems based on a systematic questionnaire. It could be shown that, especially during transfers from the operating theater to the intensive care unit or the recovery room, the time from arrival to fully centralized monitoring and the total handover time were significantly shorter with the patient-specific monitoring system (p < 0.001). Furthermore, the staff was more satisfied with the patient-specific monitor system in terms of flexibility, cleanability and usability.
Conclusion:
The increased employee satisfaction and significant time benefits during intrahospital transports may increase patient safety and efficiency of patient care, reduce employee workload, and reduce costs in the overall context of patient care.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...