Study Objective:
Nonconvulsive status epilepticus is a commonly overlooked cause of altered mental status. This study assessed nonconvulsive status epilepticus prevalence in emergency department (ED) patients with acute neurologic presentations using limited electroencephalogram (EEG) coupled with artificial intelligence (AI)-enhanced seizure detection technology. We then compared the accuracy of the AI EEG interpretations to those performed by an epileptologist.
Methods:
In a prospective observational cohort analysis, adult patients with unexplained mental status changes identified by emergency physicians received expedited placement of a limited EEG. Data collected encompassed patient demographics, clinical history, EEG interpretations by the AI algorithm and epileptologists, treatments, and disposition determinations.
Results:
There were 134 device applications on 132 patients (2 received the device twice) enrolled in the study, but 16 were missing data critical for identification or analysis and 9 did not meet the selection criteria. Of the 108 limited EEGs interpreted by an epileptologist, 69 were abnormal (diffuse slowing, highly epileptiform patterns, or spikes and sharps), 41 were normal, 5 were uninterpretable, and 3 captured episodes of seizure or status epilepticus. Limited EEG AI interpretation detected >90% seizure burden in 2 of 3 cases of seizure or status epilepticus as well as in 2 abnormal EEGs and 1 normal EEG, providing a sensitivity of 66.7% (95% confidence interval 9.4 to 99.2), a specificity of 97.0% (95% confidence interval 91.5 to 99.4), and a disease prevalence of 2.9%.
Conclusion:
Limited AI-enhanced EEG can detect nonconvulsive status epilepticus in the ED; however, the technology tended to overestimate seizure burden in our cohort. This study found a lower nonconvulsive status epilepticus prevalence compared to prior literature reports.
More Related Videos
09:00Investigating the Function of Deep Cortical and Subcortical Structures Using Stereotactic Electroencephalography: Lessons from the Anterior Cingulate Cortex
Published on: April 15, 2015
05:54Stereo-Electro-Encephalo-Graphy SEEG With Robotic Assistance in the Presurgical Evaluation of Medical Refractory Epilepsy: A Technical Note
Published on: June 13, 2016
Related Concept Videos
Epilepsy and Seizures: Overview
Various factors can trigger epilepsy, including genetic factors, brain damage, metabolic causes, and unknown etiology. Diagnosis of epilepsy involves electroencephalography (EEG), which...
Seizures l: Introduction