Related Experiment Video
Updated: Jul 1, 2026

Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
Treatment plan complexity quantification for predicting gamma passing rates in patient-specific quality assurance for
Xudong Xue1, Shunyao Luan1,2, Yi Ding1
1Department of Radiation Oncology, Hubei Cancer Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
Purpose:
To investigate the beam complexity of stereotactic Volumetric Modulated Arc Therapy (VMAT) plans quantitively and predict gamma passing rates (GPRs) using machine learning.
Methods:
The entire dataset is exclusively made of stereotactic VMAT plans (301 plans with 594 beams) from Varian Edge LINAC. The GPRs were analyzed using Varian's portal dosimetry with 2%/2 mm criteria. A total of 27 metrics were calculated to investigate the correlation between metrics and GPRs. Random forest and gradient boosting models were developed and trained to predict the GPRs based on the extracted complexity features. The threshold values of complexity metric were obtained to predict a given beam to pass or fail from ROC curve analysis.
Results:
The three moderately significant values of Spearman's rank correlation to GPRs were 0.508 (p < 0.001), 0.445 (p < 0.001), and -0.416 (p < 0.001) for proposed metric LAAM, the ratio of the average aperture area over jaw area (AAJA) and index of modulation, respectively. The random forest method achieved 98.74% prediction accuracy with mean absolute error of 1.23% using five-fold cross-validation, and 98.71% with 1.25% for gradient boosting regressor method, respectively. LAAM, leaf travelling distance (LT), AAJA, LT modulation complexity score (LTMCS) and index of modulation, were the top five most important complexity features. The LAAM metric showed the best performance with AUC value of 0.801, and threshold value of 0.365.
Conclusions:
The calculated metrics were effective in quantifying the complexity of stereotactic VMAT plans. We have demonstrated that the GPRs could be accurately predicted using machine learning methods based on extracted complexity metrics. The quantification of complexity and machine learning methods have the potential to improve stereotactic treatment planning and identify the failure of QA results promptly.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast, controlled...
Dosage Regimens: Designs and Approaches
Dosage Regimen Designs: Nomograms and Tabulations