Performance evaluation of iterative PET reconstruction with resolution recovery incorporating Gallium-68 positron
Prodromos Gavriilidis1,2,3, Michel Koole3, Anouk Marinus1,4
1Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, The Netherlands.
Background:
The distance traveled by the positron before annihilation with an electron, the so-called positron range, negatively effects the positron emission tomography (PET) image quality for radionuclides emitting high-energy positrons such as Gallium-68 (68Ga).
Purpose:
In this study, the effect of a tissue-independent positron range correction for Gallium-68 (68Ga-PRC) was investigated based on phantom measurements. The effect of the 68Ga-PRC was also explored in four patients.
Methods:
The positron range distribution profile of 68Ga in water was generated via Monte Carlo simulation. That profile was mapped to a spatially invariant 3D convolution kernel which was incorporated in the OSEM and Q.Clear reconstruction algorithms to perform the 68Ga-PRC. In addition, each reconstruction method included point spread function (PSF) modeling and time-of-flight information. For both Fluorine-18 (18F) and 68Ga, the NEMA IQ phantom was filled with a sphere-to-background ratio of 10:1 and scanned with the GE Discovery MI 5R PET/CT system. Standard non-positron range correction (PRC) reconstructions were performed for both radionuclides, while also PRC reconstructions were performed for 68Ga. Reconstructions parameters (OSEM: number of updates, Q.Clear: beta value) were adapted to achieve similar noise levels between the corresponding reconstructions. The effect of 68Ga-PRC was assessed for both OSEM and Q.Clear reconstructions and compared to non-PRC reconstructions for 68Ga and 18F in terms of image contrast, noise, recovery coefficient (RC), and spatial resolution. For the clinical validation, 68Ga-labeled prostate-specific membrane antigen (68Ga-PSMA) and 68Ga-DOTATOC PET scans were included of two patients each. For each PET scan, patients were injected with 1.5 MBq/kg of 68Ga-PSMA or 68Ga-DOTATOC and the contrast-to-noise ratio (CNR) was calculated and compared to the non-PRC reconstructions.
Results:
For OSEM reconstructions, including the 68Ga-PRC improved the RC by 9.4% (3.7%-19.3%) and spatial resolution by 21.7% (4.6 mm vs. 3.6 mm) for similar noise levels. For Q.Clear reconstructions, 68Ga-PRC modeling improved the RC by 6.7% (2.8%-10.5%) and spatial resolution by 15.3% (5.9 mm vs. 5.0 mm) while obtaining similar noise levels. In the patient data, the use of 68Ga-PRC enhanced the CNR by 13.2%.
Conclusions:
Including 68Ga-PRC in the PET reconstruction enhanced the image quality of 68Ga PET data compared to the standard non-PRC reconstructions for similar noise levels. Limited patient results also supported this improvement.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
08:55A Practical Guide for the Production and PET/CT Imaging of 68Ga-DOTATATE for Neuroendocrine Tumors in Daily Clinical Practice
Published on: April 17, 2019
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...