Correlation of plasmatic sodium determined by the laboratory and that determined by the dialysis machine
Joaquim Casals1, José Jesús Broseta1, Rosa María Fernández2
1Servicio de Nefrología y Trasplante Renal, Hospital Clínic, Barcelona, Spain.
Introduction:
Changes in plasma sodium concentration (pNa, expressed in mEq/L) are common in hemodialysis (HD) patients. Hemodialysis monitors can estimate pNa by using an internal algorithm based on ion dialysance measurements. The present study studies the accuracy of the correlation between the pNa estimated by the dialysis monitor and that measured by the biochemistry laboratory at our center.
Material And Methods:
A single-centre prospective observational study in patients on a chronic HD program with the 6008 CAREsystem monitor and standard sodium (138mmol/L) and bicarbonate (32mmol/L) prescriptions. Venous blood samples were drawn from each patient before and after each HD session to ensure inter- and intra-individual validity. The pNa was measured in the biochemistry laboratory using indirect potentiometry and simultaneously the estimated pNa by the HD monitor was recorded at the beginning and at the end of the HD session. For statistical analysis, a scatterplot was made, and Spearman's correlation quotient was calculated. In addition, the differences between both methods were represented as Bland-Altman diagrams.
Results:
The pre-dialysis pNa measured in the laboratory was 137.49±3.3, and that of the monitor, 137.96±2.91, with a correlation with R2 value of 0.683 (p<0.001). The post-dialysis pNa measured in the laboratory was 137.08±2.23, and that of the monitor was 138.87±1.88, with an R2 of 0.442 (p<0.001). On the Bland-Altman plots, the pre-dialysis pNa has a systematic error of 0.49, in favor of the monitor-estimated pNa, with a 95% confidence interval (CI) of (-3.24 to a 4.22). In the post-dialysis pNa, a systematic error of 1.79 with a 95% CI of (-1.64 to 5.22) was obtained.
Conclusion:
The correlation between the pNa estimated by Fresnius 6008 CAREsystem HD monitor and that measured by the laboratory is good, especially pre-dialysis measurements. Further studies should verify the external validity of these results.
Related Concept Videos
Regulation of Sodium and Potassium
Sodium Regulation
Sodium ions make up approximately 90% of extracellular cations, with a normal blood plasma concentration of 136–148 mEq/L. A decrease in blood volume and pressure triggers the release of renin from granular cells in the juxtaglomerular complex (JGC), primarily...
Dialysis
Roles of Electrolytes: Sodium and Potassium
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...