Measurement of arterial occlusion pressure using straight and curved blood flow restriction cuffs
Pat R Vehrs1, Ron Hager1, Nathan Dayne Richards2
1Department of Exercise Sciences, 106 SFH, Brigham Young University, Provo, Utah, USA.
Abstract:
Arterial occlusion pressure (AOP) is influenced by the characteristics of the cuff used to measure AOP. Doppler ultrasound was used to measure AOP of the brachial and superficial femoral arteries using straight and curved blood flow restriction cuffs in 21 males and 21 females. Vessel diameter and blood flow were evaluated as independent predictors of AOP. Overall, there were no significant differences in AOP when using the straight and curved cuffs in the brachial (129 mmHg vs. 128 mmHg) or superficial femoral artery (202 mmHg vs. 200 mmHg), respectively. Overall, AOP was greater (p < 0.05) in males than in females in the arm (135 mmHg, 123 mmHg) and leg (211 mmHg, 191 mmHg). Brachial (0.376 mm, 0.323 mm) and superficial femoral (0.547 mm, 0.486 mm) arteries were larger (p = 0.016) in males than in females, respectively. Systolic blood pressure (SBP) and arm circumference were predictive of brachial artery AOP, whereas SBP, diastolic blood pressure, thigh circumference, and vessel diameter were predictive of superficial femoral artery AOP. Straight and curved cuffs are efficacious in the measurement of AOP in the arm and leg. Differences in vessel size may contribute to sex differences in AOP but this requires further investigation.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: