Blood pressure measurement and assessment of arterial structure and function: an expert group position paper
Roland Asmar1, George Stergiou2, Alejandro de la Sierra3
1Foundation-Medical Research Institutes. Paris France.
Insights
This paper provides a needed practice guideline for clinicians on measuring blood pressure and assessing arterial hemodynamics. It clarifies the advantages, limitations, and clinical applications of various noninvasive methods for evaluating cardiovascular health.
Area of Science:
- Cardiovascular Science
- Medical Technology
Background:
- Accurate blood pressure (BP) measurement and arterial hemodynamics assessment are crucial for understanding cardiovascular disease and risk.
- Clinicians face numerous noninvasive methods for evaluating arterial structure and function, often with unclear technical details, benefits, and drawbacks.
Abstract:
Measuring blood pressure (BP) and investigating arterial hemodynamics are essential in understanding cardiovascular disease and assessing cardiovascular risk. Several methods are used to measure BP in the doctor's office, at home, or over 24 h under ambulatory conditions. Similarly, several noninvasive methods have been introduced for assessing arterial structure and function; these methods differ for the large arteries, the small ones, and the capillaries. Consequently, when studying arterial hemodynamics, the clinician is faced with a multitude of assessment methods whose technical details, advantages, and limitations are sometimes unclear. Moreover, the conditions and procedures for their optimal implementation, and/or the reference normality values for the parameters they yield are not always taken into sufficient consideration. Therefore, a practice guideline summarizing the main methods and their use in clinical practice is needed. This expert group position paper was developed by an international group of scientists after a two-day meeting during which each of the most used methods and techniques for blood pressure measurement and arterial function and structure evaluation were presented and discussed, focusing on their advantages, limitations, indications, normal values, and their pragmatic clinical application.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: