Related Concept Videos

Raman Spectroscopy Instrumentation: Overview01:26

Raman Spectroscopy Instrumentation: Overview

324
A conventional Raman spectrophotometer includes a laser source, a sample holding system, a wavelength selector, and a detector.
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
324
Raman Spectroscopy: Overview01:20

Raman Spectroscopy: Overview

360
The underlying principle of Raman spectroscopy is based on the interaction between light and matter, specifically molecules' inelastic scattering of photons. When a monochromatic beam of light, typically from a laser source, interacts with a sample, most scattered light has the same frequency as the incident light. This is known as Rayleigh scattering.
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
360
UV–Vis Spectrometers01:14

UV–Vis Spectrometers

1.3K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
1.3K
Atomic Absorption Spectroscopy: Lab01:21

Atomic Absorption Spectroscopy: Lab

331
For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
 Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
331
  1. Home
  2. Research Domains
  3. Engineering
  4. Communications Engineering
  5. Signal Processing
  6. Raman Spectrometer With Vertical Flow Method For Solutions Containing Organic Solvents

Raman spectrometer with vertical flow method for solutions containing organic solvents

Ting-Hao Chen1, Yu-Sheng Chen1, Hirotsugu Hiramatsu1

  • 1Department of Applied Chemistry and Institute of Molecular Science, National Yang Ming Chiao Tung University, Hsinchu 30010, Taiwan.

Spectrochimica Acta. Part A, Molecular and Biomolecular Spectroscopy
|June 21, 2024

Related Experiment Videos

Raman and IR Spectroelectrochemical Methods as Tools to Analyze Conjugated Organic Compounds
09:11

Raman and IR Spectroelectrochemical Methods as Tools to Analyze Conjugated Organic Compounds

Published on: October 12, 2018

18.3K
Real-time Monitoring of Reactions Performed Using Continuous-flow Processing: The Preparation of 3-Acetylcoumarin as an Example
09:56

Real-time Monitoring of Reactions Performed Using Continuous-flow Processing: The Preparation of 3-Acetylcoumarin as an Example

Published on: November 18, 2015

9.7K
A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer
07:52

A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer

Published on: April 12, 2017

12.8K

View abstract on PubMed

Summary
This summary is machine-generated.

This study presents a modified vertical flow (VF) Raman spectroscopy method. The enhanced VF technique and spatial mask effectively suppress solvent signals, enabling clear solute spectra detection.

Area of Science:

  • Analytical Chemistry
  • Spectroscopy
  • Physical Chemistry

Background:

  • Raman spectroscopy is a powerful technique for molecular analysis.
  • Organic solvents often exhibit strong Raman signals that can obscure solute spectra.
  • Existing methods struggle to effectively mitigate solvent interference in Raman measurements.

Purpose of the Study:

  • To develop an improved vertical flow (VF) Raman spectroscopy system.
  • To eliminate or minimize the interference from organic solvent signals.
  • To enhance the detection of solute molecules in solution.

Main Methods:

  • Construction of a Raman spectrometer utilizing the vertical flow (VF) method.
  • Incorporation of a spatial line rejection mask to filter out solvent bands.
Keywords:
Selective elimination of Raman bandsSignal enhancement methodSpatial mask

Related Experiment Videos

Raman and IR Spectroelectrochemical Methods as Tools to Analyze Conjugated Organic Compounds
09:11

Raman and IR Spectroelectrochemical Methods as Tools to Analyze Conjugated Organic Compounds

Published on: October 12, 2018

18.3K
Real-time Monitoring of Reactions Performed Using Continuous-flow Processing: The Preparation of 3-Acetylcoumarin as an Example
09:56

Real-time Monitoring of Reactions Performed Using Continuous-flow Processing: The Preparation of 3-Acetylcoumarin as an Example

Published on: November 18, 2015

9.7K
A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer
07:52

A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer

Published on: April 12, 2017

12.8K
  • Modification of the VF unit design for enhanced resistance to organic solvents.
  • Main Results:

    • A VF unit with a 60-µm pinhole achieved a 168-fold signal enhancement.
    • The spatial mask effectively removed dominant solvent Raman bands.
    • Improved dynamic range led to a 10% signal-to-noise ratio increase in methanol and acetonitrile.
    • Successful acquisition of solute spectra without solvent band interference.

    Conclusions:

    • The optimized VF Raman spectroscopy system with a spatial mask effectively suppresses solvent signals.
    • This approach significantly enhances the ability to analyze solute molecules in organic solvents.
    • The developed method offers a robust solution for sensitive Raman spectroscopic analysis in solution.