Effectiveness of the BreatheSuite Device in Assessing the Technique of Metered-Dose Inhalers: Validation Study
Meshari F Alwashmi1, Gerald Mugford2, Brett Vokey1
1BreatheSuite Inc, St. John's, NL, Canada.
Background:
The majority of medications used in treating asthma and chronic obstructive pulmonary disease (COPD) are taken through metered-dose inhalers (MDIs). Studies have reported that most patients demonstrate poor inhaler technique, which has resulted in poor disease control. Digital Health applications have the potential to improve the technique and adherence of inhaled medications.
Objective:
This study aimed to validate the effectiveness of the BreatheSuite MDI device in assessing the technique of taking a dose via an MDI.
Methods:
The study was a validation study. Thirty participants who self-reported a diagnosis of asthma or COPD were recruited from community pharmacies in Newfoundland and Labrador, Canada. Participants used a BreatheSuite MDI device attached to a placebo MDI and resembled taking 3 doses. Pharmacists used a scoring sheet to evaluate the technique of using the MDI. An independent researcher compared the results of the pharmacist's scoring sheet with the results of the BreatheSuite device.
Results:
This study found that the BreatheSuite MDI can objectively detect several errors in the MDI technique. The data recorded by the BreatheSuite MDI device showed that all participants performed at least one error in using the MDI. The BreatheSuite device captured approximately 40% (143/360) more errors compared to observation alone. The distribution of participants who performed errors in MDI steps as recorded by BreatheSuite compared to errors reported by observation alone were as follows: shaking before actuation, 33.3% (30/90) versus 25.5% (23/90); upright orientation of the inhaler during actuation, 66.7% (60/90) versus 18.87% (17/90); coordination (actuating after the start of inhalation), 76.6% (69/90) versus 35.5% (32/90); and duration of inspiration, 96.7% (87/90) versus 34.4% (31/90).
Conclusions:
The BreatheSuite MDI can objectively detect several errors in the MDI technique, which were missed by observation alone. It has the potential to enhance treatment outcomes among patients with chronic lung diseases.
Related Concept Videos
Inhaled Medications
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes:
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...