Accuracy and Functionality of Select Continuous Glucose Monitoring Systems Are Not Impacted by Implantable

Dylan Lazar1, Michael Tao2, William Matievich2

  • 1Cardiac Rhythm Management, Abbott, Sylmar, CA, USA.

Insights

Continuous glucose monitoring (CGM) systems and implantable cardioverter defibrillator (ICD) devices can safely coexist. This study found that ICDs do not interfere with the accuracy or functionality of FreeStyle Libre CGM systems during clinical use.

Area of Science:

  • Biomedical Engineering
  • Medical Device Technology
  • Endocrinology

Background:

  • Continuous glucose monitoring (CGM) is increasingly used by diabetic patients for self-management.
  • Many diabetic patients also have advanced heart disease requiring implantable cardioverter defibrillator (ICD) devices.
  • The interaction between ICDs and CGM systems is not well understood.

Purpose of the Study:

  • To evaluate the wireless coexistence and potential interference between FreeStyle Libre (FL) CGM systems and implantable cardioverter defibrillator (ICD) devices.
  • To assess the accuracy and functionality of FL CGM systems when exposed to various ICD functions and wireless interference sources.

Main Methods:

  • Tested FL CGM systems alongside ICD devices and various wireless interference sources (Wi-Fi, Bluetooth, cellular, BLE, ICD programmers).
  • Utilized five testing methodologies to assess CGM accuracy and functionality under ICD operations like high-energy shocking, pacing modes, and ATP.
  • Evaluated wireless communication and data transfer integrity between the devices.

Main Results:

  • All acceptance criteria and testing requirements for wireless coexistence were met for both CGM and ICD systems.
  • No significant interference was observed that compromised CGM accuracy or functionality.

Conclusions:

  • Coexisting ICD devices and FL CGM systems demonstrate safe and effective wireless communication.
  • Data transfer from FL CGM systems remains functional and accurate in the presence of ICDs during simulated clinical scenarios.