Three Different Protocols of Corneal Collagen Crosslinking in Keratoconus: Conventional, Accelerated and Iontophoresis
Published on: November 12, 2015
Assessment of the Predictive Ability of Theranostics for Corneal Cross-linking in Treating Keratoconus: A Randomized
Anna Maria Roszkowska1, Vincenzo Scorcia2, Rita Mencucci3
1Ophthalmology Unit, Gaetano Martino Hospital, University of Messina, Messina, Italy; Department of Ophthalmology, Andrzej Frycz Modrzewski University, Krakow, Poland.
Purpose:
To validate the ability of theranostic imaging biomarkers in assessing corneal cross-linking (CXL) efficacy in flattening the maximum keratometry (Kmax) index.
Design:
Prospective, randomized, multicenter, masked clinical trial (ClinicalTrails.gov identifier, NCT05457647).
Participants:
Fifty patients with progressive keratoconus.
Intervention:
Participants were stratified to undergo epithelium-off (25 eyes) and epithelium-on (25 eyes) CXL protocols using an ultraviolet A (UV-A) medical device with theranostic software. The device controlled UV-A light both for performing CXL and assessing the corneal riboflavin concentration (riboflavin score) and treatment effect (theranostic score). A 0.22% riboflavin formulation was applied onto the cornea for 15 minutes and 20 minutes in epithelium-off and epithelium-on protocols, respectively. All eyes underwent 9 minutes of UV-A irradiance at 10 mW/cm2.
Main Outcome Measures:
The primary outcome measure was validation of the combined use of theranostic imaging biomarkers through measurement of their accuracy (proportion of correctly classified eyes) and precision (positive predictive value) to classify eyes correctly and predict a Kmax flattening at 1 year after CXL. Other outcome measures included change in Kmax, endothelial cell density, uncorrected and corrected distance visual acuity, manifest spherical equivalent refraction and central corneal thickness 1 year after CXL.
Results:
Accuracy and precision of the theranostic imaging biomarkers in predicting eyes that had >0.1 diopter (D) of Kmax flattening at 1 year were 91% and 95%, respectively. The Kmax value significantly flattened by a median of -1.3 D (IQR, -2.11 to -0.49 D; P < 0.001); both the uncorrected and corrected distance visual acuity improved by a median of -0.1 logarithm of the minimum angle of resolution (logMAR; IQR, -0.3 to 0.0 logMAR [P < 0.001] and -0.2 to 0.0 logMAR [P < 0.001], respectively). No significant changes in endothelial cell density (P = 0.33) or central corneal thickness (P = 0.07) were noted 1 year after surgery.
Conclusions:
The study demonstrated the efficacy of integrating theranostics in a UV-A medical device for the precise and predictive treatment of keratoconus with epithelium-off and epithelium-on CXL protocols. Concentration of riboflavin and its UV-A light mediated photoactivation in the cornea are the primary factors determining CXL efficacy.
Financial Disclosure(S):
Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.
More Related Videos
05:56Scleral Cross-linking Using Riboflavin and Ultraviolet-A Radiation for Prevention of Axial Myopia in a Rabbit Model
Published on: April 3, 2016
12:25Second Harmonic Generation Signals in Rabbit Sclera As a Tool for Evaluation of Therapeutic Tissue Cross-linking TXL for Myopia
Published on: January 6, 2018
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment