Biofidelity and Limitations of Instrumented Mouthguard Systems for Assessment of Rigid Body Head Kinematics
Mitchell Z Abrams1, Jay Venkatraman2, Donald Sherman2
1Department of Biomedical Engineering, Duke University, 101 Science Dr, 1427 FCIEMAS Bldg - Box 90281, Durham, NC, 27708, USA. mitchell.abrams@duke.edu.
Abstract:
Instrumented mouthguard systems (iMGs) are commonly used to study rigid body head kinematics across a variety of athletic environments. Previous work has found good fidelity for iMGs rigidly fixed to anthropomorphic test device (ATD) headforms when compared to reference systems, but few validation studies have focused on iMG performance in human cadaver heads. Here, we examine the performance of two boil-and-bite style iMGs in helmeted cadaver heads. Three unembalmed human cadaver heads were fitted with two instrumented boil-and-bite mouthguards [Prevent Biometrics and Diversified Technical Systems (DTS)] per manufacturer instructions. Reference sensors were rigidly fixed to each specimen. Specimens were fitted with a Riddell SpeedFlex American football helmet and impacted with a rigid impactor at three velocities and locations. All impact kinematics were compared at the head center of gravity. The Prevent iMG performed comparably to the reference system up to ~ 60 g in linear acceleration, but overall had poor correlation (CCC = 0.39). Prevent iMG angular velocity and BrIC generally well correlated with the reference, while underestimating HIC and overestimating HIC duration. The DTS iMG consistently overestimated the reference across all measures, with linear acceleration error ranging from 10 to 66%, and angular acceleration errors greater than 300%. Neither iMG demonstrated consistent agreement with the reference system. While iMG validation efforts have utilized ATD testing, this study highlights the need for cadaver testing and validation of devices intended for use in-vivo, particularly when considering realistic (non-idealized) sensor-skull coupling, when accounting for interactions with the mandible and when subject-specific anatomy may affect device performance.
More Related Videos
06:52An Inertial Measurement Unit Based Method to Estimate Hip and Knee Joint Kinematics in Team Sport Athletes on the Field
Published on: May 26, 2020
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014