Temporomandibular Joint Pain Measurement by Bite Force and Von Frey Filament Assays in Mice
Published on: September 13, 2024
Laboratory Validation of a Novel Indigenously Developed Bite Force Measuring Device
Madhu Ranjan1, Surender Kumar1, Bishnupati Singh1
1Prosthodontics, Dental Institute, Rajendra Institute of Medical Sciences, Ranchi, IND.
The new BYTE device accurately measures maximum voluntary bite force in humans. Lab calibration confirms its precision and reliability for dental applications.
Area of Science:
- Biomedical Engineering
- Dental Instrumentation
Background:
- Accurate measurement of maximum voluntary bite force is crucial in restorative dentistry.
- A novel indigenous device, BYTE, has been developed for human bite force measurement.
Purpose of the Study:
- To evaluate the consistency and accuracy of the BYTE device under laboratory conditions.
Main Methods:
- The BYTE device underwent laboratory testing and calibration.
- Forces ranging from 3 N to 444 N were applied incrementally using a calibrated load cell.
- The device's recorded force values in Newtons were documented.
Main Results:
- The BYTE device demonstrated a minimum accuracy error of 0.333 N and a maximum error of 1.667 N.
- The device showed a slight underestimation of applied load, with an average accuracy error of 0.833 N.
Conclusions:
- Laboratory calibration confirms the BYTE device is precise and reliable.
- The BYTE device is suitable for measuring maximum bite force in clinical settings.