Related Experiment Video
Updated: Jun 23, 2025

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
First experience with a vendor-neutral three-dimensional mapping system for cardiac magnetic resonance-guided
Giulia De Zan1,2, Marjolein de Jongh3, Vjeran Karloci4
1Department of Thoracic, Heart and Vascular Diseases, Maggiore della Carità Hospital, Novara, Italy.
Background:
Fluoroscopy-guided catheter ablation has become the gold standard for treatment of cardiac arrhythmias. High resolution electro-anatomical mapping systems have become fundamental to perform these procedures. Recently, interventional cardiac magnetic resonance (iCMR) has been proposed as an alternative for fluoroscopy to guide atrial flutter ablations. The clinical experience with iCMR and dedicated three-dimensional mapping systems is growing. NorthStar is currently the first available vendor-neutral mapping system.
Case Summary:
We performed a real-time CMR-guided cavotricuspid isthmus (CTI) catheter ablation (CA) on a 69-year-old man using a novel mapping system (NorthStar Mapping System, Imricor Medical Systems, MN, USA). Starting from the CMR imaging, a pre-rendered segmentation model was loaded on NorthStar and used to guide the catheters, display voltage and activation maps, show mapping and ablation points. NorthStar can also take full control of the CMR scanner (i.e. start/stop sequences for anatomical information, tissue characterization, and catheter visualization) and communicate with the recorder/stimulator system (Advantage-MR EP, Imricor Medical Systems, MN, USA). With comparable procedural time to standard fluoroscopy-guided CA, CTI bidirectional block was achieved, without any complication.
Discussion:
Using the NorthStar Mapping System, we managed to achieve a successful CMR-guided CTI ablation without any complication. Its further use should be explored, especially in more complex arrhythmias where a substrate-guided ablation is critical, as it could significantly improve results in terms of arrhythmia recurrence.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
09:13Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013