Comparison of Noninvasive Oscillometric and Intra-Arterial Blood Pressure Measurements in Children Admitted to the
Jaswinder Kaur1, Siddharth Bhargava2, Puneet Aulakh Pooni2
1Division of Pediatric Gastroenterology, Hepatology & Liver Transplantation, Institute of Child Health, Sir Ganga Ram Hospital, New Delhi, India.
Abstract:
Intra-arterial blood pressure (IABP) measurement, although considered the gold standard in critically ill children, is associated with certain risks and lacks widespread availability. This study was conducted to determine the differences and agreements between oscillometric non-invasive blood pressure (NIBP) and invasive IABP measurements in children. Inclusion criteria consisted of children (from 1 month to 18 years) admitted to the pediatric intensive care unit (PICU) of a teaching hospital who required arterial catheter insertion for blood pressure (BP) monitoring. The comparison between IABP and NIBP was studied using paired t -test, Bland-Altman analysis, and Pearson's correlation coefficient. In total, 4,447 pairs of simultaneously recorded hourly NIBP and IABP measurements were collected from 65 children. Mean differences between IABP and NIBP were -3.6 ± 12.85, -4.7 ± 9.3, and -3.12 ± 9.30 mm Hg for systolic, diastolic, and mean arterial BP, respectively ( p < 0.001), with wide limits of agreement. NIBP significantly overestimated BP ( p < 0.001) in all three BP states (hypotensive, normotensive, and hypertensive), except systolic blood pressure (SBP) during hypertension where IABP was significantly higher. The difference in SBP was most pronounced during hypotension. The difference in SBP was significant in children <10 years ( p < 0.001), with the maximum difference being in infants. It was insignificant in adolescents ( p = 0.28) and underweight children ( p = 0.55). NIBP recorded significantly higher BP in all states of BP except SBP in the hypertensive state. SBP measured by NIBP tended to be the most reliable in adolescents and underweight children. NIBP was the most unreliable in infants, obese children, and during hypotension.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure